Pulmonary vein re-isolation for atrial fibrillation using duty-cycled phased radiofrequency ablation: safety and efficacy of a primary 2:1 bipolar/unipolar ablation mode.
Аннотация
PURPOSE: Pulmonary vein isolation (PVI) using phased radiofrequency (RF) energy has been shown to be effective in the treatment of paroxysmal atrial fibrillation (AF). We characterize pulmonary vein (PV) conduction at repeat ablation in patients with AF after an initial successful PVI using phased RF technology and analyze the effects of a primary 2:1 ablation mode.
METHODS AND
RESULTS
A primary 4:1 bipolar/unipolar mode in group A patients (n = 22) was compared with a primary 2:1 mode in group B (n = 22) acutely and during follow-up. Of all PVs, 81 % showed reconnection(s); 52 % of them had reconnected in all PVs. PVI was achieved in all patients without complications. Procedure and fluoroscopy times were shorter in group B (108 ± 15 vs. 126 ± 24 min and 17 ± 5 vs. 23 ± 7 min, respectively). This was attributed to a significant decrease of early PV reconnections within the first 30 min in 17 % of group B patients vs. 45 % of group A patients (p < 0.001). After 9.5 ± 4 months, recurrence of AF was detected in 5 of 22 patients (22.7 %) in group A vs. 3 of 22 patients (13.6 %) in group B (p = 0.722).
CONCLUSIONS
Phased RF energy applied by a 2:1 bipolar/unipolar mode seems safe and effective in redo-PVI procedures, resulting in a mid-term freedom from AF in 86.4%. Significant shorter procedure and fluoroscopy times compared with a primary 4:1 ablation mode during repeat PVI are mainly attributed to a lower incidence of acutely reconnected PVs within the first 30 min.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.