Dukkipati Srinivas R, Neuzil Petr, Kautzner Josef, Petru Jan, Wichterle Dan, Skoda Jan, Cihak Robert, Peichl Petr, Dello Russo Antonio, Pelargonio Gemma, Tondo Claudio, Natale Andrea, Reddy Vivek Y

Аннотация

BACKGROUND: The visually guided laser ablation (VGLA) catheter is a compliant, variable-diameter balloon that delivers laser energy around the pulmonary vein (PV) ostium under real-time endoscopic visualization. While acute PV isolation has been shown to be feasible, limited data exist regarding the durability of isolation.

OBJECTIVE: We sought to determine the durability of PV isolation following ablation using the balloon-based VGLA catheter.

METHODS: The VGLA catheter was evaluated in patients with paroxysmal atrial fibrillation (3 sites, 10 operators). Following transseptal puncture, the VGLA catheter was advanced through a 12-F deflectable sheath and inflated at the target PV ostium. Under endoscopic guidance, the 30° aiming arc was maneuvered around the PV and laser energy was delivered to ablate tissue in a contiguous/overlapping manner. At ∼3 months, all patients returned for a PV remapping procedure.

RESULTS: In 56 patients, 202 of 206 PVs (98%) were acutely isolated. At 105 ± 44 (mean ± SD) days, 52 patients returned for PV remapping at which time 162 of 189 PVs (86%) remained isolated and 32 of 52 patients (62%) had all PVs still isolated. On comparing the operators performing <10 vs ≥ 10 procedures, the durable PV isolation rate and the percentage of patients with all PVs isolated were found to be 73% vs 89% (P = .011) and 57% vs 66% (P = .746), respectively. After 2 procedures and 12.0 ± 1.9 months of follow-up, the drug-free rate of freedom from atrial fibrillation was 71.2%.

CONCLUSIONS: In this multicenter, multioperator experience, VGLA resulted in a very high rate of durable PV isolation with a clinical efficacy similar to that of radiofrequency ablation.