MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/s0091-6749(99)70440-9

Медицинская библиотека · The Journal of allergy and clinical immunology · 1999

Diagnostic performance of Food and Drug Administration-cleared serologic assays for natural rubber latex-specific IgE antibody. The Multi-Center Latex Skin Testing Study Task Force.

Hamilton R G, Biagini R E, Krieg E F

СтатьяПоддержка исследования: вне правительства СШАПоддержка службы здравоохранения СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

In the absence of Food and Drug Administration-approved natural rubber latex skin testing reagents, latex-specific IgE antibody immunoassays are used in the diagnosis of latex allergy. Comparative diagnostic performance of these tests has not been definitively determined.

OBJECTIVE

We sought to study the predictive value of available Food and Drug Administration (510K)-cleared latex-specific IgE antibody immunoassays in the diagnosis of latex allergy.

METHODS

Subjects (n = 312) were classified as having a positive (n = 117) or a negative (n = 195) latex allergy history (Hx) or having a positive (n = 131) or a negative (n = 181) puncture skin test (PST) response (Greer reagent). The 14 subjects with a negative Hx and a positive PST response had negative responses to glove provocation testing and thus were considered sensitized but asymptomatic. Sera from 22 subjects were split to evaluate intra-assay variation. All 334 coded sera were analyzed for latex-specific IgE antibodies in the Diagnostic Products Corporation microplate AlaSTAT, Hycor HY-TEC EIA System, and Pharmacia-UpJohn CAP System. Variance and diagnostic performance parameters of each test were computed with 95% confidence intervals in relation to the subjects' Hx and PST status.

RESULTS

Intra-assay concordance of split sera results was 96.0% for all 3 assays, with coefficients of variation of less than 25% and between-assay coefficients of variation of less than 21%. The diagnostic performance of the CAP and AlaSTAT assays were equivalent in comparison with PST results: sensitivity, CAP 76.3% and ALASTAT 73. 3% and specificity, CAP 96.7% and AlaSTAT 97.2% (P = NS). The HY-TEC assay was more sensitive (91.6%) and less specific (73.3%) than the CAP and AlaSTAT assays (P <.001). From 9% to 25% of the sera were discordant, being positive in at least 1, but not all 3, assays.

CONCLUSION

The CAP and AlaSTAT assays produce 24% and 27% of false-negative results, respectively, whereas the HY-TEC produces 27% of false-positive results when compared with the PST.

Для цитирования: Hamilton R G, Biagini R E, Krieg E F. Diagnostic performance of Food and Drug Administration-cleared serologic assays for natural rubber latex-specific IgE antibody. The Multi-Center Latex Skin Testing Study Task Force.. The Journal of allergy and clinical immunology. 1999. DOI: 10.1016/s0091-6749(99)70440-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.