Long-term complications in patients implanted with subcutaneous implantable cardioverter-defibrillators: Real-world data from the extended ELISIR experience.
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Gasperetti Alessio, Schiavone Marco, Ziacchi Matteo, Vogler Julia, Breitenstein Alexander, Laredo Mikael, Palmisano Pietro, Ricciardi Danilo, Mitacchione Gianfranco, Compagnucci Paolo, Bisignani Antonio, Angeletti Andrea, Casella Michela, Picarelli Francesco, Fink Thomas, Kaiser Lukas, Hakmi Samer, Calò Leonardò, Pignalberi Carlo, Santini Luca, Lavalle Carlo, Pisanò Ennio, Olivotto Iacopo, Tondo Claudio, Curnis Antonio, Dello Russo Antonio, Badenco Nicolas, Steffel Jan, Love Charles J, Tilz Roland, Forleo Giovanni, Biffi Mauro
Аннотация
OBJECTIVE: The aim of this study was to evaluate device-related complications over time in a large real-world multicenter S-ICD cohort.
METHODS: Patients implanted with an S-ICD from January 2015 to June 2020 were enrolled from a 19-institution European registry (Experience from the Long-term Italian S-ICD registry [ELISIR]; ClinicalTrials.gov identifier NCT0473876). Device-related complication rates over follow-up were collected. Last follow-up of patients was performed after the Boston Scientific recall issue.
RESULTS: A total of 1254 patients (median age 52.0 [interquartile range 41.0-62.2] years; 973 (77.6%) men; 387 (30.9%) ischemic) was enrolled. Over a follow-up of 23.2 (12.8-37.8) months, complications were observed in 117 patients (9.3%) for a total of 127 device-related complications (23.6% managed conservatively and 76.4% required reintervention). Twenty-seven patients (2.2%) had unanticipated generator replacement after 3.6 (3.3-3.9) years, while 4 (0.3%) had lead rupture. Body mass index (hazard ratio [HR] 1.063 [95% confidence interval 1.028-1.100]; P < .001), chronic kidney disease (HR 1.960 [1.191-3.225]; P = .008), and oral anticoagulation (HR 1.437 [1.010-2.045]; P = .043) were associated with an increase in overall complications, whereas older age (HR 0.980 [0.967-0.994]; P = .007) and procedure performed in high-volume centers (HR 0.463 [0.300-0.715]; P = .001) were protective factors.
CONCLUSION: The overall complication rate over 23.2 months of follow-up in a multicenter S-ICD cohort was 9.3%. Early unanticipated device battery depletions occurred in 2.2% of patients, while lead fracture was observed in 0.3%, which is in line with the expected rates reported by Boston Scientific.