Accuracy of videokeratography for instantaneous radius in keratoconus.
Аннотация
PURPOSE: Present videokeratographs provide corneal radius of curvature measurements in terms of axial (sagittal) radius and, for some instruments, instantaneous (tangential) radius. The EyeSys videokeratograph allows conversion from axial to instantaneous radius based on an algorithm, which assumes a basic corneal shape of elliptical form.
METHOD: We tested the accuracy of the EyeSys instantaneous radius algorithm for 29 keratoconic corneas using the criterion that instantaneous radius of the cone apex could be determined accurately by aligning the cone apex with the optic axis of the videokeratoscope.
RESULTS
There was a close relation between the instantaneous values obtained from the regular and apex alignments for the 17 eyes with cone apices below 55 D (95% limits of agreement -0.39 to 1.49 D), but not for the 12 eyes above 55 D (95% limits of agreement -4.87 to 6.61 D). The mean distance of the cone position from the center of the map was 1.16 mm for the instantaneous map and 1.59 mm for the axial map.
CONCLUSION
Assuming the criterion for accuracy to be the corneal radii found when the videokeratograph was aligned with the cone apex, the instantaneous map for the EyeSys videokeratograph at regular alignment provided reasonably valid measurements for corneas up to 55 D, but large errors existed for some corneas of higher power.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.