Bioprosthetic valve fracture for valve-in-valve transcatheter aortic valve implantation in patients with structural valve degeneration: Systematic review with meta-analysis.
Аннотация
OBJECTIVES
To determine the outcomes of bioprosthetic valve fracture (BVF) in valve-in-valve transcatheter aortic valve implantation (ViV-TAVI) for patients with structural valve degeneration (SVD) of bioprosthetic surgical valves (BSV) implanted during surgical aortic valve replacement (SAVR).
METHODS
A systematic review was conducted including studies published by May 2021. The primary endpoints of the study were 30-day mortality, annular rupture, stroke, paravalvular leak, pacemaker implantation, and coronary obstruction. The secondary endpoints were mean valve gradients (mmHg) and aortic valve area (AVA-cm2 ). A meta-analysis was conducted using the software R, version 3.6.3 (R Foundation for Statistical Computing).
RESULTS
Four studies including 242 patients met our eligibility criteria. The overall proportions for 30-day mortality, annular rupture, stroke, paravalvular leak, pacemaker implantation and coronary obstruction were 2.1%, <1.0%, <1.5%, <1.0%, <1.0%, and <1.5%, respectively. After ViV-TAVI with BVF, the difference in means for mean valve gradients showed a significant reduction (random-effects model: -26.7; -28.8 to -24.7; p < .001), whereas the difference in means for AVA showed a significant increase (random-effects model: 0.55 cm2 ; 0.13-0.97; p = .029). Despite the improvement in AVA means, these remain too low after the procedure highly likely due to the small size of the bioprosthetic valves implanted during the index SAVR.
CONCLUSION
ViV-TAVI with BVF has proven to be a promising option but data are still too scarce to enable us to draw definitive conclusions. Despite the decrease in gradients, postprocedural AVA remains worrisome. Studies with better designs and larger sample sizes are needed to advance this treatment option.