Cardiovascular and interventional radiology · 2024

In-Stent Restenosis in Peripheral Arterial Disease: Ultra-High-Resolution Photon-Counting Versus Third-Generation Dual-Source Energy-Integrating Detector CT Phantom Study in Seven Different Stent Types.

СтатьяСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Dachs Theresa-Marie, Hauck Sven R, Kern Maximilian, Klausenitz Catharina, Hoffner Maximilian, Schernthaner Melanie, Abdel-Rahman Hanaa, Hannover Albert, Strassl Andreas, Steiner Irene, Loewe Christian, Funovics Martin A

Аннотация

PURPOSE: The visualization of peripheral in-stent restenosis using energy-integrating detector CT is challenging due to deficient spatial resolution and artifact formation. This study compares the first clinically available photon-counting detector CT to third-generation dual-source energy-integrating detector CT.

MATERIALS AND METHODS: Nylon cylinders with central bores (4 mm, 2 mm), mimicking 75% and 95% stenoses, were placed inside seven different 8-mm diameter stents and filled with diluted contrast medium. Phantoms were scanned with photon-counting detector CT at slice thicknesses of 0.2 mm (available only in this scanner type), 0.5 mm, and 1.0 mm versus 0.5 mm and 1.0 mm in energy-integrating detector CT at matched CT dose indices. Contrast-to-noise ratios were calculated from attenuation rates. Residual lumen size was measured as full width at half-maximum. Subjective image quality was assessed by two independent blinded raters.

RESULTS: Mean contrast-to-noise ratio was lowest in photon-counting detector CT at 0.2 mm slice thickness (0%, 75%, and 95% in-stent restenosis: 6.11 ± 0.6, 5.27 ± 0.54, and 5.02 ± 0.66) and highest at 1.0 mm slice thicknesses with similar measurements in photon-counting detector CT and energy-integrating detector CT (11.46 ± 1.08, 9.94 ± 1.01, 8.26 ± 1.0 vs. 3.34 ± 1.0, 9.92 ± 0.38, 7.94 ± 1.07). Mean full width at half-maximum measurements in photon-counting detector CT at 0.2 mm slice thickness for 0%, 75%, and 95% in-stent restenosis were 8.00 ± 0.37, 3.98 ± 0.34, and 1.92 ± 0.16 mm. Full width at half-maximum was least precise in 95% in-stent restenosis at 1.0 mm slice thickness with similar measurements between scanners (1.57 ± 0.33 vs. 1.71 ± 0.15 mm). Interrater correlation coefficient was 0.75 [95% CI: [0.53; 0.86]; subjective scores were best at 0.2 mm slice thickness in photon-counting detector CT (19.43 ± 0.51 and 19.00 ± 0.68).

CONCLUSION: In phantom in-stent restenosis in 8 mm stents, we observed similar full width at half-maximum for photon-counting detector CT and energy-integrating detector CT in 0% and 75% in-stent restenosis, but at 95% in-stent restenosis, FWHM tended to be more accurate in smaller slice thicknesses in both scanners. Subjective image assessment yielded best results at 0.2 mm slice thickness in photon-counting detector CT despite lower contrast-to-noise ratio.