MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/030089160208800336

Медицинская библиотека · Tumori · 2002

New localization technique for breast cancer biopsy: mammotome guidance with imaging probe.

Soluri A, Scafè R, Falcini F, Sala R, Burgio N, Stella S, David V, Scopinaro F

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS AND

BACKGROUND

The "Imaging Probe" (IP) is a small, portable, high-resolution gamma camera to be used in radioguided surgery. The present work discusses a special prototype designed for guiding biopsies. The IP was mounted to a Fischer digital X-ray stereotactic core biopsy system in such a way that biopsy could be guided simultaneously by X-ray stereotaxis and 99mTc-Sestamibi (MIBI) images from IP.

METHODS

The IP field of view was 22.8 x 22.8 mm(2), with a spatial resolution of approx. 2.5 mm. We used off-line software for image fusion on a dedicated Pentium III portable PC. It was matched with a Fischer digital X-ray stereotactic biopsy system dedicated to direct the mammotome towards breast opacities. The operator was allowed to slightly correct the direction of the mammotome needle taking into account stereotactic X-ray, scintigraphic and fused images. Biopsy samples were counted by IP before they were sent to the pathologist.

RESULTS

High-resolution IP scintigraphy showed substantial, though not exact, matching between MIBI hot spots and X-ray opacities. More than one hot spot was detected even in the smallest (0.6 cm) lesion. Post-biopsy scintigraphy showed absence of significant hot spots in two patients, whereas in the third patient one of the three hot spots was still partially present. All lesions showed cancer on histological examination.

CONCLUSIONS

Measurement of radioactivity in biopsy specimens confirmed the heterogeneous distribution of radioactivity within cancers that IP had detected before biopsy.

Для цитирования: Soluri A, Scafè R, Falcini F, Sala R, Burgio N, Stella S, David V, Scopinaro F. New localization technique for breast cancer biopsy: mammotome guidance with imaging probe.. Tumori. 2002. DOI: 10.1177/030089160208800336.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.