Acute and long-term atrial and ventricular stimulation thresholds with a steroid-eluting electrode.

Kruse I M, Terpstra B

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Thirty leads with a steroid-eluting electrode (Medtronic 4003 and 4503) have been implanted in 24 patients, 11 in the atrium and 19 in the ventricle. Six patients received the steroid lead in both atrium and ventricle. The stimulation thresholds were followed using Elema pulse generators with Vario function in 15 patients (11 atrial leads and 10 ventricular leads) during 11 +/- 3 months (mean +/- SD). At 0.5 ms pulse duration the mean atrial and ventricular thresholds were 0.55 Volt +/- 0.22 and 0.39 Volt +/- 0.22 respectively at implant, 0.94 Volt +/- 0.13 and 0.82 Volt +/- 0.16 after 12 months. During the entire follow-up period both atrial and ventricular mean thresholds never exceeded 1 V.

Для цитирования: Kruse I M, Terpstra B. Acute and long-term atrial and ventricular stimulation thresholds with a steroid-eluting electrode.. Pacing and clinical electrophysiology : PACE. 1985. DOI: 10.1111/j.1540-8159.1985.tb05721.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.