Mapping the collateral network: Optimal near-infrared spectroscopy optode placement.

von Aspern Konstantin, Haunschild Josephina, Khachatryan Zara, Simoniuk Urszula, Ossmann Susann, Borger Michael A, Etz Christian D

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Paraplegia after extensive aortic procedures is a disastrous complication, and maintenance of adequate spinal cord perfusion/oxygenation is pivotal to its prevention. Collateral network (CN) near-infrared spectroscopy (cnNIRS) has been introduced as a noninvasive method for indirect spinal cord oxygenation monitoring. However, the CN has not been investigated in its entirety using this monitoring modality. This study aimed to identify the optimal cnNIRS positioning in an acute large animal model for routine clinical use.

METHODS

The paraspinous CN was measured from the high thoracic region to the low lumbar region (T4-L5) using cnNIRS in 10 juvenile pigs (plus reference data from 7 animals) during aortic ischemia and reperfusion. These data were compared with data on direct regional tissue perfusion of the CN and the spinal cord.

RESULTS

After aortic cross-clamping, cnNIRS at the mid-thoracic to the low lumbar level decreased rapidly to a nadir at 10 minutes of distal ischemia (mean difference, 18.3 ± 11% to 44.5 ± 9%; P < .001 to .045), with more pronounced changes in the caudal regions. High thoracic cnNIRS remained stable (mean difference, 4.3 ± 4%; P = .915). Measurements of cnNIRS, CN, and spinal cord regional perfusion demonstrated comparable curve progressions starting from the mid-thoracic region (r = 0.5-0.7; P < .001).

CONCLUSIONS

cnNIRS is capable of detecting relevant changes during ischemia and reperfusion from the mid-thoracic level downward with characteristic oxygenation patterns corresponding to CN and spinal cord regional perfusion. For extensive aortic procedures, noninvasive cnNIRS placement appears to be useful from the mid-thoracic level (T7-T9) to the lower lumbar level (L3-L5) and also may serve as a versatile monitoring method for procedures limited to the proximal thoracic aorta.

Для цитирования: von Aspern Konstantin, Haunschild Josephina, Khachatryan Zara, Simoniuk Urszula, Ossmann Susann, Borger Michael A, Etz Christian D. Mapping the collateral network: Optimal near-infrared spectroscopy optode placement.. The Journal of thoracic and cardiovascular surgery. 2020. DOI: 10.1016/j.jtcvs.2020.07.103.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.