MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/embc.2019.8857215

Медицинская библиотека · Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2019

Radio-Frequency Coil Array for Improved Concurrent Transcranial Magnetic Stimulation and Functional Magnetic Resonance Imaging.

Mathieu William, Popovich Milica, Farivar Reza

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To perform concurrent TMS-fMRI on difficult targets, such as the occipital lobe.

METHODS

a 3-channel flexible, thin RF coil was constructed that allows for whole head image coverage without impeding TMS placement. A custom MR-safe patient table which mates with typical 3T MR scanners was constructed for the purpose of face-down subject positioning to allow for access to posterior TMS targets. To counterbalance the loading effect of the TMS coil on the RF coil circuits "Dummy Loads" are introduced, which mimic the loading conditions of the TMS coil when it is not present.

RESULTS

the designed RF coil performed as expected, achieving acceptable SNR values at depths equal to the center of the average human head, and high SNR values near the surface.

CONCLUSION

The system allows for concurrent TMS-fMRI, targeting any area of the being while imaging the entire volume.

Для цитирования: Mathieu William, Popovich Milica, Farivar Reza. Radio-Frequency Coil Array for Improved Concurrent Transcranial Magnetic Stimulation and Functional Magnetic Resonance Imaging.. Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference. 2019. DOI: 10.1109/embc.2019.8857215.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.