Repeatability and Reproducibility of Slit Lamp, Optical Coherence Tomography, and Scheimpflug Measurements of Corneal Scars.
СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Das Manoranjan, Menda Shivali A, Panigrahi Arun K, Venkatesh Prajna N, Yen Michael, Tsang Betty, Kumar Alisha, Rose-Nussbaumer Jennifer, Acharya Nisha R, McCulloch Charles E, Lietman Thomas M, McLeod Stephen D, Keenan Jeremy D
Аннотация
Purpose: To determine the repeatability and reproducibility of anterior segment optical coherence tomography (AS-OCT) and Scheimpflug photography for several measurements of corneal scars, including scar size, scar depth, and corneal thickness. Methods: A series of patients treated for fungal keratitis at a tertiary eye care center in South India were recalled two years after successful treatment. Eyes with corneal scars had a slit lamp examination performed by two ophthalmologists masked to the other's examination. For AS-OCT and Scheimpflug photography, each eye had two scans taken by one technician and a third scan taken by a separate technician. Scar measurements were subsequently assessed from AS-OCT images by three graders masked to each other's results. Repeatability and reproducibility were assessed by calculating the intra-class correlation coefficient (ICC) from mixed effects linear regression models. Results: Fifty eyes had all measurements taken. The corneal scar size, measured as the geometric mean of the two longest perpendicular meridians, ranged from 0.8 to 5.4 (mean 2.8 mm, 95%CI 2.6 to 3.1). Scar size measurements taken by two separate individuals were most reproducible when the border of the scar was traced from the OCT (ICC 0.90, 95%CI 0.86 to 0.94), and least repeatable when assessed from slit lamp examination (ICC 0.80, 95%CI 0.70 to 0.90). Conclusions: AS-OCT and Scheimpflug imaging of corneal scars produced measurements with acceptable reproducibility that could be useful as cornea-specific outcomes for clinical trials.