DOI 10.1016/j.jacep.2021.10.001

Медицинская библиотека · JACC. Clinical electrophysiology · 2021

Intracardiac Echocardiography-Guided Left Atrial Appendage Closure With a Novel Foam-Based Conformable Device: Safety and 1-Year Outcomes.

Turagam Mohit K, Neuzil Petr, Hala Pavel, Mraz Tomas, Dukkipati Srinivas R, Reddy Vivek Y

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

This is a first report of the safety and 1-year outcomes of left atrial appendage closure (LAAC) using a novel foam-based conformable device, guided by intracardiac echocardiography (ICE).

BACKGROUND

Limitations of current transcatheter LAAC devices include the need for precise coaxial delivery into the left atrial appendage (LAA), potential for traumatic implantation, incomplete LAA seal, and device-related thrombus.

METHODS

The device (Conformal Left Atrial Appendage Seal, Conformal Medical Inc) is a self-expanding occluder consisting of a cylindrical nitinol endoskeleton with low-profile anchor barbs around the midpoint, covered with a porous foam cup. In a prospective single-center series, under conscious sedation, the device was delivered under fluoroscopic and ICE guidance. After positioning, a transesophageal echocardiography probe was placed to confirm ICE findings before device release. After closure, dual antiplatelet therapy was administered for 6 months. Follow-up imaging was planned for 45 days and 6 and 12 months.

RESULTS

A total of 15 patients (age 71.3 ± 10.8 years, 33% men, CHA2DS2-VASc 4.1 ± 1.7, HAS-BLED 3.4 ± 1.4) underwent LAAC, 100% successfully. There were no procedure/device-related complications requiring intervention. Asymptomatic pericardial effusion occurred in 2 patients. The 45-day, 6-month, and 12-month follow-up imaging in 11, 9, and 13 patients, respectively, revealed adequate LAA seal (leak ≤5 mm) in all patients; device-related thrombus was detected in 1 patient at 6 months. Over 1-year follow-up, there were no ischemic strokes and 1 minor bleed. Nonprocedure-/device-related death occurred in 2 patients.

CONCLUSIONS

This first report indicates that LAAC with the conformable implant guided by ICE imaging is feasible with encouraging 1-year clinical outcomes. (The Conformal Prague Study; NCT04193826).

Для цитирования: Turagam Mohit K, Neuzil Petr, Hala Pavel, Mraz Tomas, Dukkipati Srinivas R, Reddy Vivek Y. Intracardiac Echocardiography-Guided Left Atrial Appendage Closure With a Novel Foam-Based Conformable Device: Safety and 1-Year Outcomes.. JACC. Clinical electrophysiology. 2021. DOI: 10.1016/j.jacep.2021.10.001.