DOI 10.1007/s11282-024-00773-3
Accuracy of intraoral ultrasound to evaluate alveolar bone level: an ex vivo study in human cadavers.
Аннотация
OBJECTIVES
This study aimed to evaluate the reliability and accuracy of an intraoral ultrasound (US) device to evaluate alveolar bone by comparing it between different raters and to microCT (µCT) measurements.
METHODS
38 teeth distributed across three human cadavers were prepared by placing two notches on the facial enamel surface. The maxillary and mandibular teeth were imaged with a custom-designed intraoral 20 MHz ultrasound and µCT with 0.03 mm voxel size. µCT was considered the reference standard for this study. For each sample, the distance from the inferior border of the most apical notch to the tip of the alveolar bone crest on the facial aspect of the teeth was measured from the US and µCT images. Intraclass correlation coefficient (ICC) and standard deviation were calculated.
RESULTS
The intra-examiner and inter-examiner reliability for both the µCT and US alveolar bone measurements were found to be excellent (intra-examiner ICC was 0.998 for µCT and 0.997 for US, inter-examiner ICC was 0.996 for µCT and between 0.947 and 0.950 for US). The accuracy of the US was found to be good compared to µCT (ICC between 0.885 and 0.894).
CONCLUSION
The study demonstrated that intraoral ultrasound is highly reliable and accurate compared to the µCT reference standard for evaluating facial alveolar bone height.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.