Assessment of diagnostic accuracy of SanketLife - A wireless, pocket-sized ECG biosensor, in comparison to standard 12 lead ECG in the detection of cardiovascular diseases in a tertiary care setting.

Kumar Siva, Nagesh C M, Singh Manmohan, Pandian Anbu, Delurgio David, Khan Bobby, Chaudhary Robin, Gupta Prashant

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The SanketLife is a low cost, portable, pocket sized 12 lead ECG mechanised by SanketLife app running on compatible iOS and Android phones that connect wirelessly via Bluetooth technology to the device.

OBJECTIVE

The current study was conducted to assess the diagnostic accuracy of SanketLife ECG in comparison to standard 12 lead ECG (GE-2000) in detection of cardiovascular diseases.

RESEARCH DESIGN AND

METHODS

This was a prospective diagnostic test accuracy trial conducted in outpatient settings of a tertiary cardiac care centre in India. A total of 100 patients, attended cardiology OPD, were included in the study. Consecutive ECGs were taken by 12 lead standard ECG as well as by SanketLife ECG. Diagnostic accuracy variables such as sensitivity, specificity, negative and positive predictive value, negative and positive likelihood ratios were estimated. Ethical permission was taken from the Institutional ethical committee.

RESULTS &

CONCLUSION

The analysis showed a high degree of agreement and accuracy of SanketLife in detecting major cardiovascular conditions (Major Minnesota codes) such as Left and right bundle branch block, ST-segment elevation and ST-segment depression, AV conduction block. SanketLife showed high sensitivity (98.15%) and specificity (100%) in diagnosing major cardiovascular conditions.

Для цитирования: Kumar Siva, Nagesh C M, Singh Manmohan, Pandian Anbu, Delurgio David, Khan Bobby, Chaudhary Robin, Gupta Prashant. Assessment of diagnostic accuracy of SanketLife - A wireless, pocket-sized ECG biosensor, in comparison to standard 12 lead ECG in the detection of cardiovascular diseases in a tertiary care setting.. Indian pacing and electrophysiology journal. 2019. DOI: 10.1016/j.ipej.2019.12.011.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.