Garcia-Sayan Enrique, Raghunathan Deepa, Li Flora M, Dhoble Abhijeet, Sheu Richard D, Jelacic Srdjan, Reisman Mark, Smalling Richard W, Mackensen G Burkhard

Аннотация

OBJECTIVES: We present our initial experience with the fourth-generation MitraClip™ (G4) system and propose preliminary criteria for device selection.

BACKGROUND: The MitraClip™ G4 system recently underwent a "controlled release" for transcatheter edge-to-edge mitral valve repair. The four new devices include technical improvements such as controlled gripper actuation (independent leaflet capture) and continuous left atrial pressure monitoring. To date, a patient-specific device selection algorithm, and the technology's impact on procedural times and success, have not been described.

METHODS: We present an initial multi-center experience and short-term outcomes with the new system, suggest procedural and imaging considerations, and propose initial guidance for device selection.

RESULTS: Sixty-one procedures performed by three operators at two centers between November 2019 and May 2020 were analyzed. At 30-day follow-up, there were three deaths (4.9%), four neurological events (6.6%), and seven re-hospitalizations (11.5%). Fifty-nine patients achieved device and procedural success (96.7%), and there was one device-related technical issue (1.6%). Compared to the same operators utilizing the third generation MitraClip™, the G4 system resulted in a significant reduction in the median number of clips used per patient (1 IQR 1-2 vs. 2 IQR 1-3, p = .023) and a trend toward shorter device times.

CONCLUSION: Based on our initial experience, we found that the MitraClip™ G4 system is associated with high procedural success and fewer devices needed per procedure. The expanded device options may allow a more targeted approach to the myriad of pathologic presentations of mitral regurgitation. This early experience should provide a foundational opportunity for further refinement.