Flap perfusion monitoring with an attached surface probe in microvascular reconstruction of the oral cavity.

Ooms Mark, Winnand Philipp, Heitzer Marius, Peters Florian, Bock Anna, Katz Marie, Hölzle Frank, Modabber Ali

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Postoperative flap monitoring is essential in oral microvascular reconstruction for timely detection of vascular compromise. This study investigated the use of attached surface probes for the oxygen-2-see (O2C) analysis system (LEA Medizintechnik, Germany) for intraoral flap perfusion monitoring.

MATERIALS AND

METHODS

The study included 30 patients who underwent oral reconstruction with a microvascular radial-free forearm flap (RFFF) or anterolateral thigh flap (ALTF) between 2020 and 2022. Flap perfusion was measured with attached (3-mm measurement depth) and unattached surface probes (2- and 8-mm measurement depths) for the O2C analysis system at 0, 12, 24, 36, and 48 h postoperatively. Flap perfusion monitoring with attached surface probes was evaluated for cut-off values for flap blood flow, hemoglobin concentration, and hemoglobin oxygen saturation indicative of vascular compromise and for accuracy and concordance with unattached surface probes.

RESULTS

Three RFFFs were successfully revised, and one ALTF was unsuccessfully revised. The cut-off values indicative of vascular compromise for flap perfusion monitoring with attached surface probes were for RFFF and ALTF: blood flow < 60 arbitrary units (AU) and < 40AU, hemoglobin concentration > 100AU and > 80AU (both > 10% increase), and hemoglobin oxygen saturation < 40% and < 30%. Flap perfusion monitoring with attached surface probes yielded a 97.1% accuracy and a Cohen's kappa of 0.653 (p < 0.001).

CONCLUSIONS

Flap perfusion monitoring with attached surface probes for the O2C analysis system detected vascular compromise accurately and concordantly with unattached surface probes.

CLINICAL RELEVANCE: Attached surface probes for the O2C analysis system are a feasible option for intraoral flap perfusion monitoring.

Для цитирования: Ooms Mark, Winnand Philipp, Heitzer Marius, Peters Florian, Bock Anna, Katz Marie, Hölzle Frank, Modabber Ali. Flap perfusion monitoring with an attached surface probe in microvascular reconstruction of the oral cavity.. Clinical oral investigations. 2023. DOI: 10.1007/s00784-023-05177-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.