Successful percutaneous retrieval of a detached microcatheter tip using the guide-extension catheter trapping technique: A case report.

Клинический случайСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Tsujimura Takuya, Ishihara Takayuki, Iida Osamu, Asai Mitsuyoshi, Masuda Masaharu, Okamoto Shin, Nanto Kiyonori, Kanda Takashi, Okuno Shota, Matsuda Yasuhiro, Mano Toshiaki

Аннотация

A 55-year-old male with stable angina pectoris was referred to our hospital. Coronary angiography showed severe stenosis from the proximal to the middle part of the left circumflex artery (LCX). Based on the severity of lesion calcification, we performed percutaneous coronary intervention with rotational atherectomy. After crossing a 0.014-inch guidewire, we replaced it with a rotawire using a microcatheter. Following the retrieval of the microcatheter, we noted that a radiopaque tip of the microcatheter had been detached at the site of severe stenosis in the proximal part of the LCX. We advanced a guide-extension catheter to this site and wedged the detached tip using a balloon catheter (2.0-mm diameter; 12-mm length) in the guide-extension catheter. The detached tip was successfully retrieved along with the guide-extension catheter. After passing two guidewires into the main vessel and the side branch, we dilated the lesion using a 2.5-mm non-compliant balloon. Finally, we implanted two sirolimus-eluting stents, followed by post-dilatation with a 3.75-mm non-compliant balloon. This approach resulted in excellent dilatation and blood flow. Use of the guide-extension catheter trapping technique (i.e. use of a guide-extension catheter and a small balloon catheter) resulted in the successful percutaneous retrieval of a detached microcatheter tip. <Learning objective: We occasionally encounter adverse events related to the detachment or breakage of devices when performing percutaneous coronary intervention. The guide-extension catheter trapping technique using a guide-extension catheter and a small balloon is a useful approach for the retrieval of dislodged or entrapped devices.>.

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты