Freehand Three-Dimensional Ultrasound Image-Guided Needle Targeting for Percutaneous Abdominal Puncture.
Аннотация
BACKGROUND
Percutaneous abdominal interventions require precise and real-time guidance for safe needle placement. Conventional CT or fluoroscopy provides accurate localisation but involves radiation, while ultrasound (US) offers a radiation-free alternative but suffers from limited 3D visualisation.
METHODS
We developed a guidance system integrating a needle-guide mechanism with a tracked US probe to achieve real-time radiation-free targeting. The approach reconstructs freehand 3D US volumes at 25 frames per second, enabling path planning and augmenting the needle trajectory onto live B-scan images. Validation was conducted using a mockup, two phantoms, a pig model, and a human volunteer.
RESULTS
The system completed 3D US acquisition within 10 s and achieved high accuracy, with mean calibration errors of 0.62 mm and 0.22°. Needle tracking yielded angular and positional errors of 0.84 ± 0.40° and 0.76 ± 0.38 mm.
CONCLUSIONS
This method enables accurate, rapid, and radiation-free guidance, demonstrating strong potential for clinical application.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.