Diagnostic Accuracy of a Multi-Target Artificial Intelligence Service for the Simultaneous Assessment of 16 Pathological Features on Chest and Abdominal CT.

Nechaev Valentin A, Kashtanova Nataliya Y, Kopeykin Evgenii V, Magomedova Umamat M, Gribkova Maria S, Hardin Anton V, Sekacheva Marina I, Sanikovich Varvara D, Chernina Valeria Y, Gombolevskiy Victor A

СтатьяПеревод пока не подготовлен

Аннотация

Background/

Objectives

Chest, abdominal, and pelvic computed tomography (CT) with intravenous contrast is widely used for tumor staging, treatment planning, and therapy monitoring. The integration of artificial intelligence (AI) services is expected to improve diagnostic accuracy across multiple anatomical regions simultaneously. We aimed to evaluate the diagnostic accuracy of a multi-target AI service in detecting 16 pathological features on chest and abdominal CT images.

Methods

We conducted a retrospective study using anonymized CT data from an open dataset. A total of 229 CT scans were independently interpreted by four radiologists with more than 5 years of experience and analyzed by the AI service. Sixteen pathological features were assessed. AI errors were classified as minor, intermediate, or clinically significant. Diagnostic accuracy was evaluated using the area under the receiver operating characteristic curve (AUC).

Results

Across 229 CT scans, the AI service made 423 errors (11.5% of all evaluated features, n = 3664). False positives accounted for 262 cases (61.9%) and false negatives for 161 (38.1%). Most errors were minor (62.9%) or intermediate (31.7%), while clinically significant errors comprised only 5.4%. The overall AUC of the AI service was 0.88 (95% CI: 0.87-0.89), compared with 0.78-0.81 for radiologists. For clinically significant findings, the AI AUC was 0.90 (95% CI: 0.71-1.00). Diagnostic accuracy was unsatisfactory only for urolithiasis.

Conclusions

The multi-target AI service demonstrated high diagnostic accuracy for chest and abdominal CT interpretation, with most errors being clinically negligible; performance was limited for urolithiasis.

Для цитирования: Nechaev Valentin A, Kashtanova Nataliya Y, Kopeykin Evgenii V, Magomedova Umamat M, Gribkova Maria S, Hardin Anton V, Sekacheva Marina I, Sanikovich Varvara D, Chernina Valeria Y, Gombolevskiy Victor A. Diagnostic Accuracy of a Multi-Target Artificial Intelligence Service for the Simultaneous Assessment of 16 Pathological Features on Chest and Abdominal CT.. Diagnostics (Basel, Switzerland). 2025. DOI: 10.3390/diagnostics15212778.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.