Pu Ke, Shi Jing-Hong, Wang Xu, Tang Qian, Wang Xin-Jie, Tang Kai-Lin, Long Zhong-Qi, Hu Xing-Sheng

Аннотация

AIM: To investigate the diagnostic accuracy of FibroScan (FS) in detecting esophageal varices (EV) in cirrhotic patients.

METHODS: Through a systemic literature search of multiple databases, we reviewed 15 studies using endoscopy as a reference standard, with the data necessary to calculate pooled sensitivity (SEN) and specificity (SPE), positive and negative LR, diagnostic odds ratio (DOR) and area under receiver operating characteristics (AUROC). The quality of the studies was rated by the Quality Assessment of Diagnostic Accuracy studies-2 tool. Clinical utility of FS for EV was evaluated by a Fagan plot. Heterogeneity was explored using meta-regression and subgroup analysis. All statistical analyses were conducted via Stata12.0, MetaDisc1.4 and RevMan5.

RESULTS: In 15 studies (n = 2697), FS detected the presence of EV with the summary sensitivities of 84% (95%CI: 81.0%-86.0%), specificities of 62% (95%CI: 58.0%- 66.0%), a positive LR of 2.3 (95%CI: 1.81-2.94), a negative LR of 0.26 (95%CI: 0.19-0.35), a DOR of 9.33 (95%CI: 5.84-14.92) and an AUROC of 0.8262. FS diagnosed the presence of large EV with the pooled SEN of 0.78 (95%CI: 75.0%-81.0%), SPE of 0.76 (95%CI: 73.0%-78.0%), a positive and negative LR of 3.03 (95%CI: 2.38-3.86) and 0.30 (95%CI: 0.23-0.39) respectively, a summary diagnostic OR of 10.69 (95%CI: 6.81-16.78), and an AUROC of 0.8321. A meta-regression and subgroup analysis indicated different etiology could serve as a potential source of heterogeneity in the diagnosis of the presence of EV group. A Deek's funnel plot suggested a low probability for publication bias.

CONCLUSION: Using FS to measure liver stiffness cannot provide high accuracy for the size of EV due to the various cutoff and different etiologies. These limitations preclude widespread use in clinical practice at this time; therefore, the results should be interpreted cautiously given its SEN and SPE.

Предметные рубрики MeSH