DOI 10.1016/j.athoracsur.2024.02.017

Медицинская библиотека · The Annals of thoracic surgery · 2024

Preprocedural Computed Tomography Planning for Surgical Aortic Valve Replacement.

Okada Atsushi, Beckmann Erik, Rocher Erick E, Fukui Miho, Wang Cheng, Phichaphop Asa, Koike Hideki, Thao Kiahltone R, Willett Andrew, Walser-Kuntz Evan, Stanberry Larissa I, Enriquez-Sarano Maurice, Lesser John R, Sun Benjamin, Steffen Robert J, Sorajja Paul, Cavalcante João L, Bapat Vinayak N

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Selection of transcatheter valve size using preprocedural computed tomography (CT) is standardized and well established. However, valve sizing for surgical aortic valve replacement (SAVR) is currently performed intraoperatively by using sizers, which may result in variation among operators and risk for prosthesis-patient mismatch. This study evaluated the usefulness of CT annulus measurement for SAVR valve sizing.

METHODS

This study included patients who underwent SAVR using Inspiris or Magna Ease and received preoperative electrocardiogram-gated CT imaging. Starting from June 2022, study investigators applied a CT sizing algorithm using CT-derived annulus size to guide minimum SAVR label size. The final decision of valve selection was left to the operating surgeon during SAVR. The study compared the appropriateness of valve selection (comparing implanted size with CT-predicted size) and prosthesis-patient mismatch rates without aortic root enlargement between 2 cohorts: 102 cases since June 2022 (CT sizing cohort) and 180 cases from 2020 to 2021 (conventional sizing cohort).

RESULTS

Implanted size smaller than CT predicted size and severe prosthesis-patient mismatch were significantly lower by CT sizing than by conventional sizing (12% vs 31% [P = .001] and 0% vs 6% [P = .039], respectively). Interoperator variability was a factor associated with implanted size smaller than CT predicted with conventional sizing, whereas it became nonsignificant with CT sizing.

CONCLUSIONS

Applying CT sizing to SAVR led to improved valve size selection, less prosthesis-patient mismatch, and less interoperator variability. CT sizing for SAVR could also be used to predict prosthesis-patient mismatch before SAVR and identify patients who need aortic root enlargement.

Для цитирования: Okada Atsushi, Beckmann Erik, Rocher Erick E, Fukui Miho, Wang Cheng, Phichaphop Asa, Koike Hideki, Thao Kiahltone R, Willett Andrew, Walser-Kuntz Evan, Stanberry Larissa I, Enriquez-Sarano Maurice, Lesser John R, Sun Benjamin, Steffen Robert J, Sorajja Paul, Cavalcante João L, Bapat Vinayak N. Preprocedural Computed Tomography Planning for Surgical Aortic Valve Replacement.. The Annals of thoracic surgery. 2024. DOI: 10.1016/j.athoracsur.2024.02.017.