DOI 10.1093/ejcts/ezv411

Медицинская библиотека · European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2015

Early haemodynamic performance of a latest generation supra-annular aortic bioprosthesis: experience from a large single-centre series.

Deutsch Marcus-André, Prinzing Anatol, Fiegl Kathrin, Wottke Michael, Badiu Catalin C, Krane Markus, Goppel Gertrud, Bleiziffer Sabine, Guenzinger Ralf, Lange Ruediger

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The St Jude Medical (SJM) Trifecta bioprosthesis is a recently introduced stented trileaflet pericardial valve designed for supra-annular replacement of the aortic valve (AVR). We sought to evaluate the short-term clinical outcome and haemodynamic performance of the Trifecta valve after AVR.

METHODS

A total of 837 patients with severe symptomatic aortic valve stenosis or regurgitation underwent AVR with the SJM Trifecta aortic valve prosthesis between January 2009 and March 2013. All intra- and postoperative data were collected prospectively. At discharge, transthoracic echocardiography was performed. A complete set of echocardiographic data was available in 723 patients.

RESULTS

Adjusted mean systolic pressure gradients (MPGs) for valve sizes 19 (n = 37/4.4%), 21 (n = 192/22.9%), 23 (n = 263/31.4%), 25 (n = 202/24.1%), 27 (n = 100/11.9%) and 29 mm (n = 42/5.0%) were 8.6 ± 1.1, 8.7 ± 0.4, 7.2 ± 0.3, 6.2 ± 0.3, 5.6 ± 0.3 and 3.9 ± 0.4 mmHg, respectively. Mean effective orifice area (EOA) for valve sizes 19, 21, 23, 25, 27 and 29 mm were 1.5 ± 0.09, 1.6 ± 0.04, 1.9 ± 0.03, 2.0 ± 0.03, 2.2 ± 0.05 and 2.7 ± 0.01 cm(2), respectively. No patient-prosthesis mismatch (PPM) was seen in 71.3% of patients (EOAI >0.85 cm(2)/m(2)). Moderate mismatch (EOAI 0.65-0.85 cm(2)/m(2)) was observed in 23.9% of patients, whereas severe PPM (EOAI <0.65 cm(2)/m(2)) occurred in 4.4% of patients. No malfunction of the prosthesis, endocarditis, valve thrombosis or relevant aortic regurgitation necessitating surgical revision was observed until discharge.

CONCLUSIONS

The SJM Trifecta valve reveals an excellent early haemodynamic performance with low residual MPGs and a low incidence of PPM. Studies with longitudinal clinical and echocardiographic assessments with longer term follow-up evaluation including a comparison with other contemporary bioprostheses are needed.

Для цитирования: Deutsch Marcus-André, Prinzing Anatol, Fiegl Kathrin, Wottke Michael, Badiu Catalin C, Krane Markus, Goppel Gertrud, Bleiziffer Sabine, Guenzinger Ralf, Lange Ruediger. Early haemodynamic performance of a latest generation supra-annular aortic bioprosthesis: experience from a large single-centre series.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2015. DOI: 10.1093/ejcts/ezv411.