DOI 10.1016/j.rec.2015.08.010

Медицинская библиотека · Revista espanola de cardiologia (English ed.) · 2015

Differences in Frame Geometry Between Balloon-expandable and Self-expanding Transcatheter Heart Valves and Association With Aortic Regurgitation.

Rodríguez-Olivares Ramón, Rahhab Zouhair, Faquir Nahid El, Ren Ben, Geleijnse Marcel, Bruining Nico, van Mieghem Nicolas M, Schultz Carl, Lauritsch Guenter, de Jaegere Peter P T

СтатьяНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION AND

OBJECTIVES

Patient- and procedure-related factors are known to be associated with aortic regurgitation after transcatheter aortic valve implantation. Nevertheless, this entity may also be caused by a specific device-host interaction due to the biomechanical properties of the valves, independently of clinical factors. We sought to elucidate the role of frame geometry in the occurrence of aortic regurgitation after Medtronic CoreValve and Edwards SAPIEN valve implantation.

METHODS

We conducted an observational study encompassing 134 patients undergoing transcatheter aortic valve implantation with the Medtronic CoreValve and Edwards SAPIEN valve. Frame analysis was performed at 3 predefined levels of both valves by rotational angiography using dedicated motion compensation software. A distinction was made between patients with no-to-mild and moderate-to-severe aortic regurgitation by echocardiography.

RESULTS

Baseline characteristics were similar between the 2 valves. Despite greater use of predilation in the CoreValve (95.2% vs 82.0%; P=.012), more oversizing (perimeter, 114±7% vs 103±7%; P<.001), and the same depth of implantation (noncoronary sinus, 7±4 vs 8±2mm; left coronary sinus, 8±4 vs 8±2mm), it was less expanded and more eccentric than the Edwards SAPIEN (83±7% vs 92±4%; P<.001 and 82±8% vs 95±3%; P<.001, respectively) and when eccentricity was adjusted for the patient's annulus eccentricity (4±13% vs 21±11%; P<.001). Eccentricity and adjusted eccentricity were associated with moderate-to-severe aortic regurgitation.

CONCLUSIONS

Independently of patient- and procedure-related factors, there is a device-specific device-host interaction that explains aortic regurgitation after transcatheter aortic valve implantation.

Для цитирования: Rodríguez-Olivares Ramón, Rahhab Zouhair, Faquir Nahid El, Ren Ben, Geleijnse Marcel, Bruining Nico, van Mieghem Nicolas M, Schultz Carl, Lauritsch Guenter, de Jaegere Peter P T. Differences in Frame Geometry Between Balloon-expandable and Self-expanding Transcatheter Heart Valves and Association With Aortic Regurgitation.. Revista espanola de cardiologia (English ed.). 2015. DOI: 10.1016/j.rec.2015.08.010.