MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2015.08.022

Медицинская библиотека · JACC. Cardiovascular interventions · 2015

Transfemoral Implantation of a Fully Repositionable and Retrievable Transcatheter Valve for Noncalcified Pure Aortic Regurgitation.

Schofer Joachim, Nietlispach Fabian, Bijuklic Klaudija, Colombo Antonio, Gatto Fernando, De Marco Federico, Mangieri Antonio, Hansen Lorenz, Bruschi Giuseppe, Ruparelia Neil, Rieß Friedrich-Christian, Maisano Franscesco, Latib Azeem

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study sought to evaluate the use of the Direct Flow Medical (DFM) transcatheter heart valve (Direct Flow Medical, Santa Rosa, California) for the treatment of noncalcific pure aortic regurgitation (AR).

BACKGROUND

The treatment of noncalcific AR has remained a relative contraindication with transcatheter heart valves due to challenges in anchoring devices in the absence of calcium, concerns of valve embolization, and the high risk of significant residual paravalvular leak.

METHODS

The study population consisted of patients treated for severe noncalcific pure AR with transfemoral implantation of a DFM transcatheter heart valve at 6 European centers. The primary endpoint was the composite endpoint of device success and the secondary endpoint was the composite early safety endpoint (according to the VARC-2 criteria).

RESULTS

Eleven high-risk (STS score 8.84 ± 8.9, Logistic EuroSCORE 19.9 ± 7.1) patients (mean age 74.7 ± 12.9 years) were included. Device success was achieved in all patients. In 1 patient, the initial valve prosthesis was retrieved after pull-through, and a second valve was successfully deployed. The early safety endpoint was reached in 91% of the patients, with 1 patient requiring surgical aortic valve replacement secondary to downward dislocation of the prosthesis that was successfully managed with surgical aortic valve replacement. DFM implantation resulted in excellent hemodynamics with none or trivial paravalvular regurgitation in 9 patients and a transprosthetic gradient of 7.7 ± 5.1 mm Hg at 30-day follow up. All patients derived symptomatic benefit following the procedure, with 72% in New York Heart Association functional class I or II.

CONCLUSIONS

This study reports the feasibility of treating severe noncalcific AR with the Direct Flow prosthesis via the transfemoral route.

Для цитирования: Schofer Joachim, Nietlispach Fabian, Bijuklic Klaudija, Colombo Antonio, Gatto Fernando, De Marco Federico, Mangieri Antonio, Hansen Lorenz, Bruschi Giuseppe, Ruparelia Neil, Rieß Friedrich-Christian, Maisano Franscesco, Latib Azeem. Transfemoral Implantation of a Fully Repositionable and Retrievable Transcatheter Valve for Noncalcified Pure Aortic Regurgitation.. JACC. Cardiovascular interventions. 2015. DOI: 10.1016/j.jcin.2015.08.022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.