DOI 10.1111/joic.12248

Медицинская библиотека · Journal of interventional cardiology · 2015

Early Outcomes With Direct Flow Medical Versus First-Generation Transcatheter Aortic Valve Devices: A Single-Center Propensity-Matched Analysis.

Giustino Gennaro, Latib Azeem, Panoulas Vasileios F, Montorfano Matteo, Chieffo Alaide, Taramasso Maurizio, Sato Katsumasa, Agricola Eustachio, Alfieri Ottavio, Colombo Antonio

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To compare early outcomes between Direct Flow Medical (DFM) and 1(st) -generation transcatheter aortic valve replacement (TAVR) devices.

BACKGROUND

The effectiveness of new-generation TAVR devices compared with earlier generations is unclear.

METHODS AND

RESULTS

Overall, 496 patients treated between November 2007 and March 2014 were included: 44 (8.7%) treated with DFM, 179 (36.9%) with Medtronic CoreValve (MCV) and 273 (54.4%) with Edwards Sapien / Sapien XT (SXT) valve. Because of differences in baseline characteristics, propensity-score matching among patients treated in the same time span (March 2012-March 2014) was performed. The primary objective was the VARC-2 device success composite endpoint. Propensity-score matching identified 41 patients in each group. A higher device success composite endpoint was observed in DFM (DFM 97.6% vs. MCV 65.9% vs. SXT 92.7%; P < 0.001). This was attributed to a significantly lower incidence of moderate-to-severe post-procedural aortic regurgitation (AR; 2.4% vs. 22% vs. 7.3%; P < 0.001), lower rate of valve embolization (0% vs. 7.3% vs. 0%; 0.041) and need for a 2nd valve implantation (0% vs. 7.3% vs. 0%; P = 0.041).

CONCLUSIONS

DFM was associated with higher rates of device success and a lower incidence of post-procedural AR. New generation devices appear to address several of the limitations of earlier generation devices.

Для цитирования: Giustino Gennaro, Latib Azeem, Panoulas Vasileios F, Montorfano Matteo, Chieffo Alaide, Taramasso Maurizio, Sato Katsumasa, Agricola Eustachio, Alfieri Ottavio, Colombo Antonio. Early Outcomes With Direct Flow Medical Versus First-Generation Transcatheter Aortic Valve Devices: A Single-Center Propensity-Matched Analysis.. Journal of interventional cardiology. 2015. DOI: 10.1111/joic.12248.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.