Does frame geometry play a role in aortic regurgitation after Medtronic CoreValve implantation?

Rodríguez-Olivares Ramón, El Faquir Nahid, Rahhab Zouhair, Geeve Patrick, Maugenest Anne-Marie, van Weenen Sander, Ren Ben, Galema Tjebbe, Geleijnse Marcel, Van Mieghem Nicolas M, van Domburg Ron, Bruining Nico, Schultz Carl, Lauritsch Guenter, de Jaegere Peter P T

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Aortic regurgitation (AR) after Medtronic CoreValve System (MCS) implantation may be explained by patient-, operator- and procedure-related factors. We sought to explore if frame geometry, as a result of a specific device-host interaction, contributes to AR.

METHODS AND

RESULTS

Using rotational angiography with dedicated motion compensation, we assessed valve frame geometry in 84 patients who underwent TAVI with the MCS. Aortic regurgitation was assessed by angiography (n=84, Sellers) and echocardiography at discharge (n=72, VARC-2). Twenty-two patients (26%) had AR grade ≥2 using contrast angiography, and 17 (24%) by echocardiography. Balloon predilatation and sizing and depth of implantation did not differ between the two groups. Despite more frequent balloon post-dilatation in patients with AR (40.9 vs. 9.7%, p=0.001), the frame was more elliptical at its nadir relative to the patient's annulus (6±13 vs. -1±11%, p=0.046) and occurred in a larger proportion of patients (61.9 vs. 26.8%, p=0.004). Although the Agatston score and the eccentricity of the MCS frame relative to the annulus were independent determinants of AR (odds ratio: 1.635 [1.151-2.324], p=0.006, and 4.204 [1.237-14.290], p=0.021), there was a weak association between the Agatston score and the adjusted eccentricity (Spearman's rank correlation coefficient =-0.24, p=0.046).

CONCLUSIONS

These findings indicate that AR can be explained by a specific device-host interaction which can only partially be explained by the calcium load of the aortic root.

Для цитирования: Rodríguez-Olivares Ramón, El Faquir Nahid, Rahhab Zouhair, Geeve Patrick, Maugenest Anne-Marie, van Weenen Sander, Ren Ben, Galema Tjebbe, Geleijnse Marcel, Van Mieghem Nicolas M, van Domburg Ron, Bruining Nico, Schultz Carl, Lauritsch Guenter, de Jaegere Peter P T. Does frame geometry play a role in aortic regurgitation after Medtronic CoreValve implantation?. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2016. DOI: 10.4244/eijy15m08_02.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.