DOI 10.1093/ehjci/jev174

Медицинская библиотека · European heart journal. Cardiovascular Imaging · 2015

Device landing zone calcification and its impact on residual regurgitation after transcatheter aortic valve implantation with different devices.

Seiffert Moritz, Fujita Buntaro, Avanesov Maxim, Lunau Clemens, Schön Gerhard, Conradi Lenard, Prashovikj Emir, Scholtz Smita, Börgermann Jochen, Scholtz Werner, Schäfer Ulrich, Lund Gunnar, Ensminger Stephan, Treede Hendrik

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: Calcification of the device landing zone is linked to paravalvular regurgitation after transcatheter aortic valve implantation (TAVI). The mechanisms remain incompletely understood and the performance of next-generation transcatheter heart valves (THV) has not been investigated. We evaluated the impact of calcification patterns on residual aortic regurgitation (AR) after TAVI with different THV in patients with severe aortic stenosis.

METHODS AND

RESULTS

TAVI was performed in 537 patients at two centres. Devices implanted were the Edwards Sapien XT (n = 254), Medtronic CoreValve (n = 123), JenaValve (n = 62), Medtronic Engager (n = 56), and Symetis Acurate (n = 42) prostheses. Calcification of the device landing zone was retrospectively assessed from contrast-enhanced multidetector computed tomography data and correlated with echocardiographic and clinical data. Calcium volumes of the aortic valve and left-ventricular outflow tract were associated with residual AR: No AR, 604 mm(3) (inter-quartile range, IQR 349-916); trace AR, 639 mm(3) (IQR 368-948); mild AR, 710 mm(3) (IQR 412-2078); ≥moderate AR, 1041 mm(3) (IQR 791-1417, P = 0.001). Device landing zone calcium, particularly if located in the left-ventricular outflow tract, and a low cover index were predictive of AR. Differences in the incidence of AR were observed with regard to THV type. Higher calcium volume was associated with the need for post-dilation (n = 134, median 852 [IQR 342-945] vs. 604 [IQR 542-1207] mm(3), P < 0.001).

CONCLUSION

Calcification of the device landing zone, particularly if located inferior to the annulus, was independently associated with residual AR after TAVI with all evaluated THV; however, the incidence of paravalvular leakage differed significantly between the devices implanted.

Для цитирования: Seiffert Moritz, Fujita Buntaro, Avanesov Maxim, Lunau Clemens, Schön Gerhard, Conradi Lenard, Prashovikj Emir, Scholtz Smita, Börgermann Jochen, Scholtz Werner, Schäfer Ulrich, Lund Gunnar, Ensminger Stephan, Treede Hendrik. Device landing zone calcification and its impact on residual regurgitation after transcatheter aortic valve implantation with different devices.. European heart journal. Cardiovascular Imaging. 2015. DOI: 10.1093/ehjci/jev174.