Transcatheter, inflatable, and fully repositionable aortic valve: Preliminary results using a modified implantation technique.

Kische Stephan, D'Ancona Giuseppe, Agma Hüseyin U, Gürer Hakan, Ortak Jasmin, Elsässer Albrecht, Öner Alper, Ince Hüseyin

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

We present our experience with an inflatable, non-metallic, fully retrievable, and repositionable transcatheter aortic valve [Direct Flow® Medical (DFM)] focusing on technical features adopted during implantation.

BACKGROUND

Implantation techniques of new generation percutaneous aortic valves are still developing and with experience implementations are described.

METHODS

The "low pressure" (initial valve inflation at 4 ATM, lower than the recommended 12 ATM) and the "alternate aortic curve" techniques (initial valve pulling either from the wire running on the side of the inner or from the one on the outer aortic curve) are part of our modified protocol for DFM® implantation.

RESULTS

Forty-two consecutive patients underwent TAVI with DFM. The "low pressure" technique was used in all patients. In 27 (64.3%) patients the bulkiest calcifications were on the inner aortic curve side, and an inner curve technique was used to initiate valve pulling toward the annular hinge point. In the remaining 15 (35.7%) patients, an outer curve technique was used. Neither pull-through nor re-valving was reported. Valve performance showed mean gradient of 8.9 mm Hg. Mild paravalvular leak was reported in seven (16%) patients. No one experienced moderate or severe aortic insufficiency. Thirty-day mortality was 9.5% (4/42).

CONCLUSION

The "low pressure technique" and the "alternate curve technique" guarantee an anatomy and patient tailored approach to achieve controlled and finely tuned valve seating.

Для цитирования: Kische Stephan, D'Ancona Giuseppe, Agma Hüseyin U, Gürer Hakan, Ortak Jasmin, Elsässer Albrecht, Öner Alper, Ince Hüseyin. Transcatheter, inflatable, and fully repositionable aortic valve: Preliminary results using a modified implantation technique.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2015. DOI: 10.1002/ccd.26039.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.