Real world outcomes using 20 mm balloon expandable SAPIEN 3/ultra valves compared to larger valves (23, 26, and 29 mm)-a propensity matched analysis.
Аннотация
OBJECTIVE/
BACKGROUND
Small balloon expandable valves have higher echocardiographic transvalvular gradients and rates of prosthesis-patient mismatch (PPM) compared to larger valves. However, the impact of these echocardiographic findings on clinical outcomes is unknown. We sought to determine the clinical outcomes of 20 mm SAPIEN 3 (S3 BEV) compared to larger S3 BEV in relation to echocardiographic hemodynamics.
METHODS
Using the STS/ACC transcatheter valve registry, we performed a propensity-matched comparison of patients undergoing treatment of native aortic valve stenosis using transfemoral, balloon-expandable implantation of 20 mm and ≥ 23 mm S3 BEVs. Baseline and procedure characteristics, echocardiographic variables and survival were analyzed. Multivariable logistic regression was used to identify predictors of 1-year mortality.
RESULTS
After propensity matching of the 20 mm and ≥ 23 mm SAPIEN 3 valves, 3,931 pairs with comparable baseline characteristics were identified. Small valves were associated with significantly higher echocardiographic gradients at discharge (15.7 ± 7.1 mmHg vs. 11.7 ± 5.5 mmHg, p < 0.0001) and severe PPM rates (21.5% vs. 9.7%, p < 0.0001). There was no significant difference in 1-year all-cause mortality (20 mm: 13.0% vs. ≥23 mm: 12.7%, p = 0.72) or other major adverse event rates and outcomes between the two cohorts. Based on a multivariable analysis, elevated discharge mean gradient (>20 mmHg), severe PPM and the use of 20 mm versus ≥23 mm were not independent predictors of 1-year mortality.
CONCLUSION
SAPIEN 3 20 mm valves were associated with higher echocardiographic gradients, and severe PPM rates compared to larger valves but these factors were not associated with significant differences in 1-year all-cause mortality or rehospitalization.