Repeat transcatheter aortic valve implantation and implications for transcatheter heart valve performance: insights from bench testing.

Sathananthan Janarthanan, Fraser Rob, Landes Uri, Rich Courtney, Sellers Stephanie L, Leipsic Jonathon, Blanke Philipp, Lutter Georg, Frank Derk, Puehler Thomas, Wood David A, Søndergaard Lars, Webb John G

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

THV implantation within failed surgical valves is well established. However, the implications of THV implantation within failed THVs are poorly understood.

AIMS: This study aimed to assess the impact of different transcatheter heart valve (THV) designs and implant positioning strategies on hydrodynamic performance after redo transcatheter aortic valve implantation (TAVI).

METHODS

THVs of varying design (SAPIEN 3, Evolut PRO, ACURATE neo, ALLEGRA, and Portico) and size were implanted inside SAPIEN XT and Evolut R THVs. Hydrodynamic function as per International Organization for Standardization (ISO) specifications was evaluated using a pulse duplicator, and multi-modality imaging was performed.

RESULTS

The majority of tested THV-in-THV combinations resulted in stable anchoring of the new implant. However, some combinations resulted in unstable anchoring with the potential for dislodgement or embolisation. Hydrodynamic function was favourable for all tested THV designs implanted in the intra-annular SAPIEN XT THV. SAPIEN 3 implantation within an Evolut THV with supra-annular leaflets must be appropriately sized to ensure adequate anchoring. Avoidance of an intra-annular deployment mitigated leaflet overhang of the Evolut leaflets and optimised hydrodynamic performance.

CONCLUSIONS

This study demonstrates that most THV designs and implantation strategies can result in favourable hydrodynamic performance following redo TAVI. Whether the leaflets of a failed THV are intra- or supra-annular may have important implications for the positioning of a redo-THV implant. Certain THV designs or implantation positions may be more desirable when performing redo TAVI.

Для цитирования: Sathananthan Janarthanan, Fraser Rob, Landes Uri, Rich Courtney, Sellers Stephanie L, Leipsic Jonathon, Blanke Philipp, Lutter Georg, Frank Derk, Puehler Thomas, Wood David A, Søndergaard Lars, Webb John G. Repeat transcatheter aortic valve implantation and implications for transcatheter heart valve performance: insights from bench testing.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2021. DOI: 10.4244/eij-d-20-00697.