DOI 10.1097/01.ico.0000159733.37554.ba

Медицинская библиотека · Cornea · 2005

Central corneal thickness measurements using Orbscan II scanning slit topography, noncontact specular microscopy, and ultrasonic pachymetry in eyes with keratoconus.

Kawana Keisuke, Miyata Kazunori, Tokunaga Tadatoshi, Kiuchi Takahiro, Hiraoka Takahiro, Oshika Tetsuro

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To compare corneal thickness measurements using Orbscan II scanning slit topography, Topcon SP-2000P noncontact specular microscopy, and ultrasonic pachymetry in eyes with keratoconus.

METHODS

Central corneal thickness was measured in 22 eyes with keratoconus. Eyes with apparent corneal opacity were excluded. Scanning slit topography, noncontact specular microscopy, and ultrasonic pachymetry were used in this sequence. The acoustic equivalent correlation factor (0.92) was used for Orbscan readings.

RESULTS

Three devices gave significantly different corneal thickness readings (P < 0.001, repeated-measure analysis of variance). Measurements with Orbscan scanning slit topography (449.5 +/- 43.2 [SD] mum) were significantly smaller than those of ultrasonic pachymetry (485.0 +/- 29.3 microm; P < 0.001, Tukey multiple comparison) and SP-2000P noncontact specular microscopy (476.7 +/- 28.3 microm; P = 0.002). There were significant linear correlations between ultrasonic pachymetry and scanning slit topography (Pearson correlation coefficient r = 0.741, P < 0.001), between scanning slit topography and noncontact specular microscopy (r = 0.880, P < 0.001), and between noncontact specular microscopy and ultrasonic pachymetry (r = 0.811, P < 0.001).

CONCLUSION

In eyes with keratoconus, Orbscan II scanning slit topography system gave significantly smaller corneal thickness readings than the other 2 devices. Measurements taken by noncontact specular microscopy and ultrasonic pachymetry were comparable. Three devices showed significant linear correlations with one another.

Для цитирования: Kawana Keisuke, Miyata Kazunori, Tokunaga Tadatoshi, Kiuchi Takahiro, Hiraoka Takahiro, Oshika Tetsuro. Central corneal thickness measurements using Orbscan II scanning slit topography, noncontact specular microscopy, and ultrasonic pachymetry in eyes with keratoconus.. Cornea. 2005. DOI: 10.1097/01.ico.0000159733.37554.ba.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.