MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00520-011-1145-5

Медицинская библиотека · Supportive care in cancer : official journal of the Multinational Association of Supportive Care in Cancer · 2011

Frequent AV node reentrant tachycardia induced by oversized port catheter.

Lange Stefan Andreas, Jung Jens

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Here, we describe the efficiency of interventional electrophysiology therapy in a critically ill patient with heart failure because of incessant AV node reentrant tachycardia after surgical port implantation.

METHODS

Electrophysiology diagnostic and radio frequency ablation were utilized in the study.

RESULT: We presented a 78-year-old patient with colorectal cancer, surgical port implantation, and progressive heart failure due to an incessant AV node reentrant tachycardia (AVNRT). This rhythm disturbance was refractory to any conventional pharmaceutical treatment. The electrophysiology diagnostic obtained an AVNRT. During fluoroscopy, an oversized port catheter with a loop touching the tricuspid valve annulus was discovered. This port catheter was responsible for premature beats, which induced incessant AVNRT. A manual reposition of the port edge and additional AV node slow pathway modification terminated further tachycardia.

CONCLUSION

Electrophysiology diagnostic and radiofrequency ablation procedures are promising techniques also in critically ill patients with hemodynamic compromising supraventricular tachycardia.

Для цитирования: Lange Stefan Andreas, Jung Jens. Frequent AV node reentrant tachycardia induced by oversized port catheter.. Supportive care in cancer : official journal of the Multinational Association of Supportive Care in Cancer. 2011. DOI: 10.1007/s00520-011-1145-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…