DOI 10.1016/j.jcin.2010.06.012

Медицинская библиотека · JACC. Cardiovascular interventions · 2010

Patent foramen ovale closure using a bioabsorbable closure device: safety and efficacy at 6-month follow-up.

Van den Branden Ben J, Post Martijn C, Plokker Herbert W, ten Berg Jurriën M, Suttorp Maarten J

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to assess the mid-term safety and efficacy of percutaneous patent foramen ovale (PFO) closure using a bioabsorbable device (BioSTAR, NMT Medical, Boston, Massachusetts).

BACKGROUND

Closure of PFO in patients with cryptogenic stroke has proven to be safe and effective using different types of permanent devices.

METHODS

All consecutive patients who underwent percutaneous PFO closure with the bioabsorbable closure device between November 2007 and January 2009 were included. Residual shunt was assessed using contrast transthoracic echocardiography.

RESULTS

Sixty-two patients (55% women, mean age 47.7 ± 11.8 years) underwent PFO closure. The in-hospital complications were a surgical device retrieval in 2 patients (3.2%), device reposition in 1 (1.6%), and a minimal groin hematoma in 6 patients (9.7%). The short-term complications at 1-month follow-up (n = 60) were a transient ischemic attack in the presence of a residual shunt in 1 patient and new supraventricular tachycardia in 7 patients (11.3%). At 6-month follow-up (n = 60), 1 patient without residual shunt developed a transient ischemic attack and 1 developed atrial fibrillation. A mild or moderate residual shunt was noted in 51.7%, 33.9%, and 23.7% after 1-day, 1-month, and 6-month follow-up, respectively. A large shunt was present in 8.3%, 3.4%, and 0% after 1-day, 1-month, and 6-month follow-up.

CONCLUSIONS

Closure of PFO using the bioabsorbable device is associated with a low complication rate and a low recurrence rate of embolic events. However, a relatively high percentage of mild or moderate residual shunting is still present at 6-month follow-up.

Для цитирования: Van den Branden Ben J, Post Martijn C, Plokker Herbert W, ten Berg Jurriën M, Suttorp Maarten J. Patent foramen ovale closure using a bioabsorbable closure device: safety and efficacy at 6-month follow-up.. JACC. Cardiovascular interventions. 2010. DOI: 10.1016/j.jcin.2010.06.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.