DOI 10.1016/j.radonc.2023.109910

Медицинская библиотека · Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2023

Stereotactic arrhythmia radioablation: A multicenter pre-post intervention safety evaluation of the implantable cardioverter-defibrillator function.

van der Ree Martijn H, Hoeksema Wiert F, Luca Adrian, Visser Jorrit, Balgobind Brian V, Zumbrink Michiel, Spier Raymond, Herrera-Siklody Claudia, Lee Justin, Bates Matthew, Daniel Jim, Peedell Clive, Boda-Heggemann Judit, Rudic Boris, Merten Roland, Dieleman Edith M, Rinaldi Cristopher A, Ahmad Shahreen, Whitaker John, Bhagirath Pranav, Hatton Matthew Q, Riley Stephen, Grehn Melanie, Schiappacasse Luis, Blanck Oliver, Hohmann Stephan, Pruvot Etienne, Postema Pieter G

Многоцентровое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Stereotactic arrhythmia radioablation (STAR) appears to be beneficial in selected patients with therapy-refractory ventricular tachycardia (VT). However, high-dose radiotherapy used for STAR-treatment may affect functioning of the patients' implantable cardioverter defibrillator (ICD) by direct effects of radiation on ICD components or cardiac tissue. Currently, the effect of STAR on ICD functioning remains unknown.

METHODS

A retrospective pre-post multicenter study evaluating ICD functioning in the 12-month before and after STAR was performed. Patients with (non)ischemic cardiomyopathies with therapy-refractory VT and ICD who underwent STAR were included and the occurrence of ICD-related adverse events was collected. Evaluated ICD parameters included sensing, capture threshold and impedance. A linear mixed-effects model was used to investigate the association between STAR, radiotherapy dose and changes in lead parameters over time.

RESULTS

In total, 43 patients (88% male) were included in this study. All patients had an ICD with an additional right atrial lead in 34 (79%) and a ventricular lead in 17 (40%) patients. Median ICD-generator dose was 0.1 Gy and lead tip dose ranged from 0-32 Gy. In one patient (2%), a reset occurred during treatment, but otherwise, STAR and radiotherapy dose were not associated with clinically relevant alterations in ICD leads parameters.

CONCLUSIONS

STAR treatment did not result in major ICD malfunction. Only one radiotherapy related adverse event occurred during the study follow-up without patient harm. No clinically relevant alterations in ICD functioning were observed after STAR in any of the leads. With the reported doses STAR appears to be safe.

Для цитирования: van der Ree Martijn H, Hoeksema Wiert F, Luca Adrian, Visser Jorrit, Balgobind Brian V, Zumbrink Michiel, Spier Raymond, Herrera-Siklody Claudia, Lee Justin, Bates Matthew, Daniel Jim, Peedell Clive, Boda-Heggemann Judit, Rudic Boris, Merten Roland, Dieleman Edith M, Rinaldi Cristopher A, Ahmad Shahreen, Whitaker John, Bhagirath Pranav, Hatton Matthew Q, Riley Stephen, Grehn Melanie, Schiappacasse Luis, Blanck Oliver, Hohmann Stephan, Pruvot Etienne, Postema Pieter G. Stereotactic arrhythmia radioablation: A multicenter pre-post intervention safety evaluation of the implantable cardioverter-defibrillator function.. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 2023. DOI: 10.1016/j.radonc.2023.109910.