Effects of the Surface Texture and Weight of a Pinch Apparatus on the Reliability and Validity of a Hand Sensorimotor Control Assessment.

Hsu Hsiu-Yun, Kuan Ta-Shen, Yang Hsiu-Ching, Tsai Ching-Liang, Yeh Chien-Hsien, Lin Chih-Chun, Kuo Li-Chieh

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

To investigate the reliability and validity of a modified pinch apparatus devised with 3 surface textures and 2 different weights for clinical application.

DESIGN

Case-controlled study.

SETTING

A university hospital.

PARTICIPANTS

The participants (N=32) included carpal tunnel syndrome (CTS) patients (n=16) with 20 sensory neuropathy hands, and an equal number of age-sex matched volunteers without CTS, as well as young volunteers without CTS (n=16 with 20 hands) used to analyze both the testing validity and reliability of the modified device.

INTERVENTIONS: Not applicable.

MAIN OUTCOME MEASURES: The Semmes-Weinstein monofilament (SWM) and two-point discrimination (2PD) tests were conducted, and the force ratio between the FPpeak (peak pinch force during lifting phase) and FLmax (maximum load force at maximum upward acceleration onset) detected from a pinch-holding-up activity (PHUA) under various testing conditions was obtained.

RESULTS

The range of the intraclass correlation coefficient of this pinch device was 0.369-0.952. The CTS patients exhibited poorer force modulation ability according to the inertial change in a dynamic lifting task when compared to the controls under all testing conditions (P<.001). The area under the receiver operating characteristic force ratio curve was 0.841, revealing high accuracy of the test for diagnosing CTS neuropathic hands under the testing condition in which the 125-g coarse texture device was used. In addition, the weight factor was shown to have significant effects on the sensitivity and accuracy of the PHUA assessment.

CONCLUSIONS

This study showed that the PHUA test via the modified pinch apparatus is a sensitive tool that can be used in clinical practice for detecting neuropathic CTS hands. In addition, changing the weight of the pinch device has a significant effect on the sensitivity and accuracy of the PHUA assessment.

Для цитирования: Hsu Hsiu-Yun, Kuan Ta-Shen, Yang Hsiu-Ching, Tsai Ching-Liang, Yeh Chien-Hsien, Lin Chih-Chun, Kuo Li-Chieh. Effects of the Surface Texture and Weight of a Pinch Apparatus on the Reliability and Validity of a Hand Sensorimotor Control Assessment.. Archives of physical medicine and rehabilitation. 2018. DOI: 10.1016/j.apmr.2018.07.440.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.