MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1542-474x.2007.00180.x

Медицинская библиотека · Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc · 2007

Automatic p wave analysis over 24 hours in patients with paroxysmal or persistent atrial fibrillation.

Bitzen Alexander, Sternickel Karsten, Lewalter Thorsten, Schwab Jörg Otto, Yang Alexander, Schrickel Jan Wilko, Linhart Markus, Wolpert Christian, Jung Werner, David Peter, Lüderitz Berndt, Nickenig Georg, Lickfett Lars

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Patients with atrial fibrillation (AF) often exhibit abnormalities of P wave morphology during sinus rhythm. We examined a novel method for automatic P wave analysis in the 24-hour-Holter-ECG of 60 patients with paroxysmal or persistent AF and 12 healthy subjects.

METHODS

Recorded ECG signals were transferred to the analysis program where 5-10 P and R waves were manually marked. A wavelet transform performed a time-frequency decomposition to train neural networks. Afterwards, the detected P waves were described using a Gauss function optimized to fit the individual morphology and providing amplitude and duration at half P wave height.

RESULTS

>96% of P waves were detected, 47.4 +/- 20.7% successfully analyzed afterwards. In the patient population, the mean amplitude was 0.073 +/- 0.028 mV (mean variance 0.020 +/- 0.008 mV(2)), the mean duration at half height 23.5 +/- 2.7 ms (mean variance 4.2 +/- 1.6 ms(2)). In the control group, the mean amplitude (0.105 +/- 0.020 ms) was significantly higher (P < 0.0005), the mean variance of duration at half height (2.9 +/- 0.6 ms(2)) significantly lower (P < 0.0085).

CONCLUSIONS

This method shows promise for identification of triggering factors of AF.

Для цитирования: Bitzen Alexander, Sternickel Karsten, Lewalter Thorsten, Schwab Jörg Otto, Yang Alexander, Schrickel Jan Wilko, Linhart Markus, Wolpert Christian, Jung Werner, David Peter, Lüderitz Berndt, Nickenig Georg, Lickfett Lars. Automatic p wave analysis over 24 hours in patients with paroxysmal or persistent atrial fibrillation.. Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc. 2007. DOI: 10.1111/j.1542-474x.2007.00180.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.