4-Pyridoxic Acid in the Spent Dialysate: Contribution to Fluorescence and Optical Monitoring.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Kalle Sigrid, Tanner Risto, Arund Jürgen, Tomson Ruth, Luman Merike, Fridolin Ivo
Аннотация
AIM: In this work we estimated the contribution of the fluorescence of 4-pyridoxic acid (4-PA) to the total fluorescence of spent dialysate with the aim of evaluating the on-line monitoring of removal of this vitamin B-6 metabolite from the blood of patients with end-stage renal disease (ESRD).
METHODS: Spectrofluorometric analysis of spent dialysate, collected from hemodialysis and hemodiafiltration sessions of 10 patients receiving regularly pyridoxine injections after dialysis treatment, was performed in the range of Ex/Em 220-500 nm. 4-PA in dialysate samples was identified and quantified using HPLC with fluorescent and MS/MS detection.
RESULTS: Averaged HPLC chromatogram of spent dialysate had many peaks in the wavelength region of Ex320/Em430 nm where 4-PA was the highest peak with contribution of 42.2±17.0% at the beginning and 47.7±18.0% in the end of the dialysis. High correlation (R = 0.88-0.95) between 4-PA concentration and fluorescence intensity of spent dialysate was found in the region of Ex310-330/Em415-500 nm, respectively.
CONCLUSION: 4-PA elimination from the blood of ESRD patients can be potentially followed using monitoring of the fluorescence of the spent dialysate during dialysis treatments.
METHODS: Spectrofluorometric analysis of spent dialysate, collected from hemodialysis and hemodiafiltration sessions of 10 patients receiving regularly pyridoxine injections after dialysis treatment, was performed in the range of Ex/Em 220-500 nm. 4-PA in dialysate samples was identified and quantified using HPLC with fluorescent and MS/MS detection.
RESULTS: Averaged HPLC chromatogram of spent dialysate had many peaks in the wavelength region of Ex320/Em430 nm where 4-PA was the highest peak with contribution of 42.2±17.0% at the beginning and 47.7±18.0% in the end of the dialysis. High correlation (R = 0.88-0.95) between 4-PA concentration and fluorescence intensity of spent dialysate was found in the region of Ex310-330/Em415-500 nm, respectively.
CONCLUSION: 4-PA elimination from the blood of ESRD patients can be potentially followed using monitoring of the fluorescence of the spent dialysate during dialysis treatments.