Effect of three-dimensional valve shape on the hemodynamics of aortic stenosis: three-dimensional echocardiographic stereolithography and patient studies.

Gilon Dan, Cape Edward G, Handschumacher Mark D, Song Jae Kwan, Solheim Joan, VanAuker Michael, King Mary Etta E, Levine Robert A

СтатьяПоддержка исследования: вне правительства СШАПоддержка службы здравоохранения СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study tested the hypothesis that the impact of a stenotic aortic valve depends not only on the cross-sectional area of its limiting orifice but also on three-dimensional (3D) valve geometry.

BACKGROUND

Valve shape can potentially affect the hemodynamic impact of aortic stenosis by altering the ratio of effective to anatomic orifice area (the coefficient of orifice contraction [Cc]). For a given flow rate and anatomic area, a lower Cc increases velocity and pressure gradient. This effect has been recognized in mitral stenosis but assumed to be absent in aortic stenosis (constant Cc of 1 in the Gorlin equation).

METHODS

In order to study this effect with actual valve shapes in patients, 3D echocardiography was used to reconstruct a typical spectrum of stenotic aortic valve geometrics from doming to flat. Three different shapes were reproduced as actual models by stereolithography (computerized laser polymerization) with orifice areas of 0.5, 0.75, and 1.0 cm(2) (total of nine valves) and studied with physiologic flows. To determine whether valve shape actually influences hemodynamics in the clinical setting, we also related Cc (= continuity/planimeter areas) to stenotic aortic valve shape in 35 patients with high-quality echocardiograms.

RESULTS

In the patient-derived 3D models, Cc varied prominently with valve shape, and was largest for long, tapered domes that allow more gradual flow convergence compared with more steeply converging flat valves (0.85 to 0.90 vs. 0.71 to 0.76). These variations translated into differences of up to 40% in pressure drop for the same anatomic area and flow rate, with corresponding variations in Gorlin (effective) area relative to anatomic values. In patients, Cc was significantly lower for flat versus doming bicuspid valves (0.73 +/- 0.14 vs. 0.94 +/- 0.14, p < 0.0001) with 40 +/- 5% higher gradients (p < 0.0001).

CONCLUSIONS

Three-dimensional valve shape is an important determinant of pressure loss in patients with aortic stenosis, with smaller effective areas and higher pressure gradients for flatter valves. This effect can translate into clinically important differences between planimeter and effective valve areas (continuity or Gorlin). Therefore, valve shape provides additional information beyond the planimeter orifice area in determining the impact of valvular aortic stenosis on patient hemodynamics.

Для цитирования: Gilon Dan, Cape Edward G, Handschumacher Mark D, Song Jae Kwan, Solheim Joan, VanAuker Michael, King Mary Etta E, Levine Robert A. Effect of three-dimensional valve shape on the hemodynamics of aortic stenosis: three-dimensional echocardiographic stereolithography and patient studies.. Journal of the American College of Cardiology. 2002. DOI: 10.1016/s0735-1097(02)02269-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.