Minute ventilation-sensor driven pacemaker related difficulty in weaning from mechanical ventilation: a case report.

Nayak A K, Aggarwal K, Flaker G C

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Minute ventilation responsive rate adaptive pacemaker improves exercise performance in the patients with chronotropic insufficiency. The usual complications of these pacemakers are related to pacemaker implantation, rather than to the rate adaptive function.

CASE REPORT: We report a patient with mitral stenosis and atrial fibrillation with a VVIR pacemaker, who was mechanically ventilated for respiratory failure. The patient could not be weaned because of the minute ventilation driven increased heart rates. Subsequently the rate adaptive function was switched off and the patient was successfully weaned from ventilator.

CONCLUSIONS

Difficulty in weaning from mechanical ventilation can be a serious complication in patients with organic heart disease, who depend on diastolic filling time for adequate cardiac output.

Для цитирования: Nayak A K, Aggarwal K, Flaker G C. Minute ventilation-sensor driven pacemaker related difficulty in weaning from mechanical ventilation: a case report.. Missouri medicine. 2001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.