DOI 10.1002/jeo2.12110

Медицинская библиотека · Journal of experimental orthopaedics · 2024

Superior load-to-failure in an all-suture anchor system for all-inside meniscal repair compared to a PEEK-cage anchor system in an experimental cadaveric test setting.

Pichler Lorenz, Kiss Gyula, Sator Thomas, Schuller Andrea, Kandathil Sam A, Hofbauer Marcus, Koch Thomas, Hirtler Lena, Tiefenboeck Thomas

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The purpose of this study was to compare the biomechanical properties of a latest generation all-suture anchor repair device (ASARD) for meniscal repair with that of a latest generation PEEK-cage anchor repair device (PCARD) in an experimental setting using cadaveric menisci.

METHODS

Twenty-six menisci were obtained from the knees of fresh body donors. Artificially created meniscal lesions were treated randomly, using a single stitch with either an ASARD or a PCARD. Cyclic biomechanical testing, utilising a universal material testing machine and following an established protocol, was carried out and load-to-failure (LTF), displacement, stiffness, and mode-of-failure (MOF) reported.

RESULTS

Mean LTF was found to be 61% higher in the ASARD group at 107.10 N (standard deviation [SD], 42.34), compared to 65.86 N (SD, 27.42) in the PCARD group with statistical significance (p = 0.022). The ASARD exhibited a trend towards higher stiffness (10.35 N; SD, 3.92 versus 7.78 N; SD; 3.59) and higher displacement at cycles one, 100, and 499 (1.64, 3.27, and 4.17 mm versus 0.93, 2.19, and 2.83 mm) compared to the PCARD. Cheese wiring was the most common mode-of-failure in both groups (76.9%).

CONCLUSIONS

This study demonstrates that an ASARD shows a higher mean LTF than a PCARD when compared in an experimental biomechanical setting.

LEVEL OF EVIDENCE: Level III.

Для цитирования: Pichler Lorenz, Kiss Gyula, Sator Thomas, Schuller Andrea, Kandathil Sam A, Hofbauer Marcus, Koch Thomas, Hirtler Lena, Tiefenboeck Thomas. Superior load-to-failure in an all-suture anchor system for all-inside meniscal repair compared to a PEEK-cage anchor system in an experimental cadaveric test setting.. Journal of experimental orthopaedics. 2024. DOI: 10.1002/jeo2.12110.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.