DOI 10.1016/j.asmr.2024.100938

Медицинская библиотека · Arthroscopy, sports medicine, and rehabilitation · 2024

Biocomposite Anchors Have Greater Yield Load and Energy Compared With All-Suture Anchors in an In Vitro Ovine Infraspinatus Tendon Repair Model.

Walton Robert A, Liuzza Lindsey, Takawira Catherine, Leonardi Claudia, Lopez Mandi J

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare tensile fatigue and strength measures of biocomposite and all-suture anchors in an ovine humerus-infraspinatus tendon model of rotator cuff repair.

METHODS

Infraspinatus tendons on adult ovine humeri were sharply transected at the insertion. One of each pair was assigned randomly for fixation with 2 biocomposite or all-suture anchors. Constructs were tested with 200 cycles of 20 to 70 N tensile load, and gap formation was measured at the incised tendon end every 50 cycles. They were subsequently tested to failure. Outcome measures including fatigue stiffness, hysteresis, creep, and gap formation and tensile stiffness, and yield and failure displacement, load, and energy were compared between anchors.

RESULTS

Biocomposite anchors had greater yield load (134.1 ± 6.5 N, P < .01) and energy (228.6 ± 85.7 J, P < .03) than all-suture anchors (104.7 ± 6.5 N, 169.8 ± 85.7 J). Fatigue properties were not different between anchors, but stiffness and gap formation increased and hysteresis and creep decreased significantly with increasing cycle number.

CONCLUSIONS

Although the yield displacement of both anchors was within the range of clinical failure, the tensile yield load and energy of ovine infraspinatus tendons secured to the humerus with 2 single-loaded all-suture anchors in a single row were significantly lower than those secured with 2 biocomposite anchors in the same configuration.

CLINICAL RELEVANCE: It is important to understand the biomechanical properties for selecting anchors for rotator cuff repair. A direct comparison of fatigue testing followed by failure strength of infraspinatus tendon fixation with all-suture and biocomposite anchors could help guide anchor selection and postoperative mobility recommendations.

Для цитирования: Walton Robert A, Liuzza Lindsey, Takawira Catherine, Leonardi Claudia, Lopez Mandi J. Biocomposite Anchors Have Greater Yield Load and Energy Compared With All-Suture Anchors in an In Vitro Ovine Infraspinatus Tendon Repair Model.. Arthroscopy, sports medicine, and rehabilitation. 2024. DOI: 10.1016/j.asmr.2024.100938.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.