Reliability of single-lead electrocardiogram interpretation to detect atrial fibrillation: insights from the SAFER feasibility study.

Hibbitt Katie, Brimicombe James, Cowie Martin R, Dymond Andrew, Freedman Ben, Griffin Simon J, Hobbs F D R Ichard, Lindén Hannah Clair, Lip Gregory Y H, Mant Jonathan, McManus Richard J, Pandiaraja Madhumitha, Williams Kate, Charlton Peter H

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Single-lead electrocardiograms (ECGs) can be recorded using widely available devices such as smartwatches and handheld ECG recorders. Such devices have been approved for atrial fibrillation (AF) detection. However, little evidence exists on the reliability of single-lead ECG interpretation. We aimed to assess the level of agreement on detection of AF by independent cardiologists interpreting single-lead ECGs and to identify factors influencing agreement.

METHODS AND

RESULTS

In a population-based AF screening study, adults aged ≥65 years old recorded four single-lead ECGs per day for 1-4 weeks using a handheld ECG recorder. Electrocardiograms showing signs of possible AF were identified by a nurse, aided by an automated algorithm. These were reviewed by two independent cardiologists who assigned participant- and ECG-level diagnoses. Inter-rater reliability of AF diagnosis was calculated using linear weighted Cohen's kappa (κw). Out of 2141 participants and 162 515 ECGs, only 1843 ECGs from 185 participants were reviewed by both cardiologists. Agreement was moderate: κw = 0.48 (95% confidence interval, 0.37-0.58) at participant level and κw = 0.58 (0.53-0.62) at ECG level. At participant level, agreement was associated with the number of adequate-quality ECGs recorded, with higher agreement in participants who recorded at least 67 adequate-quality ECGs. At ECG level, agreement was associated with ECG quality and whether ECGs exhibited algorithm-identified possible AF.

CONCLUSION

Inter-rater reliability of AF diagnosis from single-lead ECGs was found to be moderate in older adults. Strategies to improve reliability might include participant and cardiologist training and designing AF detection programmes to obtain sufficient ECGs for reliable diagnoses.

Для цитирования: Hibbitt Katie, Brimicombe James, Cowie Martin R, Dymond Andrew, Freedman Ben, Griffin Simon J, Hobbs F D R Ichard, Lindén Hannah Clair, Lip Gregory Y H, Mant Jonathan, McManus Richard J, Pandiaraja Madhumitha, Williams Kate, Charlton Peter H. Reliability of single-lead electrocardiogram interpretation to detect atrial fibrillation: insights from the SAFER feasibility study.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2024. DOI: 10.1093/europace/euae181.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.