MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/euw297

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2017

Intermuscular technique for implantation of the subcutaneous implantable cardioverter defibrillator: long-term performance and complications.

Winter Joachim, Siekiera Markus, Shin Dong-In, Meyer Christian, Kröpil Patric, Clahsen Harald, O'Connor Stephen

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: The subcutaneous cardioverter defibrillator was designed to overcome electrode complications of transvenous defibrillation systems. While largely achieved, pocket complications have increased. Subcutaneous implantation of the pulse generator leaves it prone to erosion, extrusion, discomfort, and poor cosmesis.

METHODS AND

RESULTS

We use a demonstration electrode and pulse generator with fluoroscopy, prior to prepping and draping, to maximize the left ventricular mass between them. We adapted a submuscular abdominal ICD technique to implant the S-ICD intermuscularly between the anterior surface of serratus anterior and the posterior surface of latissimus dorsi. Surgery in our patients beyond the subcutaneous tissue was bloodless, as muscle layers were carefully separated but not incised, which also protected the long thoracic nerve. Two layers of muscle protect the pulse generator. We have implanted 82 consecutive patients with this technique, taking ∼65 min. All patients were converted with 65 J standard polarity shock during induced arrhythmia conversion testing, with six (7.3%) patients requiring a repositioning of the pulse generator prior to successful conversion. Seven spontaneous episodes of ventricular fibrillation were detected in three (3.6%) patients, all successfully converted back to sinus rhythm. Long-term patient outcomes have been good with low complication rates over the mean ± standard deviation 3.6 ± 1.2 years.

CONCLUSION

Our intermuscular technique and implant methodology is successful for placement of the subcutaneous defibrillator pulse generator. Our technique leads to an excellent cosmetic result and high levels of patient satisfaction. Rates of first shock conversion during defibrillation testing, inappropriate shocks, and complications during follow-up compare favourably with previous published case series. There were no left arm movement limitations post-operatively.

Для цитирования: Winter Joachim, Siekiera Markus, Shin Dong-In, Meyer Christian, Kröpil Patric, Clahsen Harald, O'Connor Stephen. Intermuscular technique for implantation of the subcutaneous implantable cardioverter defibrillator: long-term performance and complications.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2017. DOI: 10.1093/europace/euw297.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.