Intermuscular technique for implantation of the subcutaneous implantable cardioverter defibrillator: long-term performance and complications.
Аннотация
AIMS: The subcutaneous cardioverter defibrillator was designed to overcome electrode complications of transvenous defibrillation systems. While largely achieved, pocket complications have increased. Subcutaneous implantation of the pulse generator leaves it prone to erosion, extrusion, discomfort, and poor cosmesis.
METHODS AND
RESULTS
We use a demonstration electrode and pulse generator with fluoroscopy, prior to prepping and draping, to maximize the left ventricular mass between them. We adapted a submuscular abdominal ICD technique to implant the S-ICD intermuscularly between the anterior surface of serratus anterior and the posterior surface of latissimus dorsi. Surgery in our patients beyond the subcutaneous tissue was bloodless, as muscle layers were carefully separated but not incised, which also protected the long thoracic nerve. Two layers of muscle protect the pulse generator. We have implanted 82 consecutive patients with this technique, taking ∼65 min. All patients were converted with 65 J standard polarity shock during induced arrhythmia conversion testing, with six (7.3%) patients requiring a repositioning of the pulse generator prior to successful conversion. Seven spontaneous episodes of ventricular fibrillation were detected in three (3.6%) patients, all successfully converted back to sinus rhythm. Long-term patient outcomes have been good with low complication rates over the mean ± standard deviation 3.6 ± 1.2 years.
CONCLUSION
Our intermuscular technique and implant methodology is successful for placement of the subcutaneous defibrillator pulse generator. Our technique leads to an excellent cosmetic result and high levels of patient satisfaction. Rates of first shock conversion during defibrillation testing, inappropriate shocks, and complications during follow-up compare favourably with previous published case series. There were no left arm movement limitations post-operatively.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.