MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.29785

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2021

Outflow graft obstruction in patients with the HM 3 LVAD: A percutaneous approach.

Milwidsky Assi, Alvarez Villela Miguel, Wiley Jose, Sanina Cristina, Patel Snehal R, Sutton Nicole, Latib Azeem, Sims Daniel B, Forest Stephen J, Shin Julia J, Farooq Muhammad U, Goldstein Daniel J, Jorde Ulrich P

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The use of the HeartMate 3 (HM3) left ventricular assist device (LVAD) is expanding. Despite being associated with lower rates of adverse events and increased survival, outflow graft obstruction (OGO) has been reported in patients with HM3. The incidence and best management of this serious complication remain unclear.

METHODS

We describe six cases of HM3 OGO occurring in five patients in our institutional HM3 cohort. Four cases underwent computed tomography angiography and in two percutaneous angiography was directly performed to confirm the diagnosis. In four cases, percutaneous repair of the OG was performed using common interventional cardiology (IC) techniques.

RESULTS

Our institutional incidence of OGO was 7% (event rate of 0.05 per patient year); much higher than the previously reported incidence of 1.6%. All cases occurred in the bend relief covered segment. Only two patients had apparent OG twisting, and in two, OGO occurred despite placement of an anti-twist clip at the time of implant. External compression seems to play a role in most cases. Balloon "graftoplasty" and stent deployment via the femoral artery alleviated the obstruction and normalized LVAD flow in all patients who underwent percutaneous repair. The use of self-expanding stents allowed for downsizing of the procedural access site to 10 Fr. No serious procedure-related complications occurred.

CONCLUSION

OGO is common in HM3 patients, external compression due to biomaterial accumulated surrounding the OG is a common etiology. Percutaneous repair using standard IC techniques is safe and feasible in cases of compression with or without partial twisting.

Для цитирования: Milwidsky Assi, Alvarez Villela Miguel, Wiley Jose, Sanina Cristina, Patel Snehal R, Sutton Nicole, Latib Azeem, Sims Daniel B, Forest Stephen J, Shin Julia J, Farooq Muhammad U, Goldstein Daniel J, Jorde Ulrich P. Outflow graft obstruction in patients with the HM 3 LVAD: A percutaneous approach.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2021. DOI: 10.1002/ccd.29785.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.