Comparison of Accuracy of Freestyle Libre Pro and Medtronic iPro2 Continuous Glucose Monitoring Systems in Pregnancy.

Golla Kiran Kumar, Gupta Yashdeep, Goyal Alpesh, Kalaivani Mani, Kachhawa Garima, Kulshrestha Vidushi, Sharma Aparna K, Meena Jyoti, Bharti Juhi, Sharma Jai B, Dadhwal Vatsla, Malhotra Neena, Bhatla Neerja, Tandon Nikhil

Наблюдательное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

There are limited data on head-to-head performance of Freestyle Libre Pro (FSL-Pro) and blinded Medtronic iPro2 continuous glucose monitoring system in pregnancy. In this prospective observational study, women with hyperglycemia in pregnancy (n = 42) underwent simultaneous FSL-Pro and Medtronic iPro2 sensor insertion and self-monitoring of blood glucose using Contour Plus meter (reference). The overall mean absolute relative difference (MARD) for iPro2 and FSL-Pro systems were 8.0% ± 9.2% and 19.0% ± 12.7%, respectively. At hypoglycemic range, both sensors performed less accurately (MARD: 18.0% and 16.8%, respectively), whereas iPro2 showed higher accuracy at euglycemic (8.2% and 19.3%, respectively) and hyperglycemic (6.8% and 18.0%, respectively) ranges. On Bland-Altman analysis, iPro2 and FSL-Pro underestimated glucose by 0.01 and 1.09 mmol/L, respectively. The ISO criteria were fulfilled for 88.5% and 44.9% of all values, respectively. To conclude, iPro2 was more accurate; however, both sensors demonstrated inaccuracy at hypoglycemic range, highlighting the need for refinements in the current generation of sensors to address this problem.

Для цитирования: Golla Kiran Kumar, Gupta Yashdeep, Goyal Alpesh, Kalaivani Mani, Kachhawa Garima, Kulshrestha Vidushi, Sharma Aparna K, Meena Jyoti, Bharti Juhi, Sharma Jai B, Dadhwal Vatsla, Malhotra Neena, Bhatla Neerja, Tandon Nikhil. Comparison of Accuracy of Freestyle Libre Pro and Medtronic iPro2 Continuous Glucose Monitoring Systems in Pregnancy.. Diabetes technology & therapeutics. 2023. DOI: 10.1089/dia.2023.0070.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.