DOI 10.1515/cclm-2023-0072

Медицинская библиотека · Clinical chemistry and laboratory medicine · 2023

Analytical evaluation of a GAD65 antibodies chemiluminescence immunoassay for CSF in neurological syndromes.

Musso Giulia, Zoccarato Marco, Gallo Nicoletta, Padoan Andrea, Cosma Chiara, Zuliani Luigi, De Gaspari Piera, Pegoraro Elena, Plebani Mario, Basso Daniela

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Antibodies against glutamic acid decarboxylase isoform 65 (GAD-Ab) have been found in different severe neurological conditions associated with altered synthesis of γ-aminobutyric acid (GABA). Serum GAD-Ab can be found in up to 90 % of patients with type 1 diabetes mellitus (T1DM), mostly at relatively low concentrations, while high concentrations of GAD-ab are thought to be more frequently associate to a neurological condition, with levels 100-folds higher than those found in T1DM. Although CSF testing is recommended when suspecting a GAD-associated neurological syndrome, no commercial immunoassay is validated for this use and no cut-off is internationally recognized to support the diagnosis.

METHODS

In this study we validated CSF testing of GAD-Ab on an automated chemiluminescence (CLIA) immunoassay that had previously shown good agreement with ELISA on serum.

RESULTS

We tested 43 CSF from patients with typical GAD-associated neurological disorders and patients with other neurological conditions, identifying a clinical cut-off of 18 kIU/L that discriminated GAD-disease with an area under the curve (AUC) of 0.921. CLIA showed good analytical performances on repeatability and recovery tests in CSF and confirmed an excellent agreement with ELISA.

CONCLUSIONS

GAD-Ab associated neurological disorders are rare but CSF testing for GAD-Ab is a common request for neurologists when suspecting an insidious autoimmune central nervous system disease. CLIA platforms are expected to be increasingly adopted in clinical laboratories due to their flexibility and reliability, therefore studies on decisional levels should be implemented for improving the interpretation and utilization of laboratory data.

Для цитирования: Musso Giulia, Zoccarato Marco, Gallo Nicoletta, Padoan Andrea, Cosma Chiara, Zuliani Luigi, De Gaspari Piera, Pegoraro Elena, Plebani Mario, Basso Daniela. Analytical evaluation of a GAD65 antibodies chemiluminescence immunoassay for CSF in neurological syndromes.. Clinical chemistry and laboratory medicine. 2023. DOI: 10.1515/cclm-2023-0072.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.