Development of a flow cytometry-based pulse-width assay for detection of aggregates in cellular therapeutics to be infused by catheter.

Hickerson Darren H M, White Hannah Storrie, Nguyen Alan, Nieman Elizabeth, Meadows Norin, Gentry Tracy, Foster Sandra J, Balber Andrew E, Fiordalisi Michelle, Hinson James M

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AIMS: Delivery of cell-based therapies through the carotid artery with the use of an intra-arterial catheter could introduce aggregates and cause focal ischemia in the brain. We developed a pulse-width flow cytometry method for aggregate detection and quantification. The assay was designed to be used as a cell product release assay in a clinical trial seeking to treat ischemic stroke with sorted cells brightly expressing aldehyde dehydrogenase (ALDH(br) cells) delivered through intra-arterial catheters.

METHODS

The forward light scatter pulse-width axis of a flow cytometer was calibrated for particle diameter measurements through the use of traceable standard microspheres and linear regression. As a positive control, Concanavalin A-aggregated cells were counted manually and sorted onto slides to compare with pulse width-determined values. Known numbers of aggregates were spiked into purified singlet cells for quantification. A clinical standard for aggregate count and diameter was determined. The assay was used to qualify catheters with the use of ALDH(br) cells.

RESULTS

The pulse-width axis was highly linear for microsphere diameter (r(2) > 0.99), which allowed for size calibration. Microscopically determined counts and diameters corresponded to pulse width-determined values. Known aggregate counts were linear with pulse width-determined aggregate counts (r(2) = 0.98). The limit of detection was determined to be 0.004%. Flow of ALDH(br) cells through catheters did not generate aggregates. The final method to be used as a release assay for the stroke clinical trial was tested successfully on samples from volunteer donors.

CONCLUSIONS

The pulse-width aggregate detection assay provides a reliable, reproducible, accurate and rapid means of detection, classification and quantification of aggregates in cell therapy products.

Для цитирования: Hickerson Darren H M, White Hannah Storrie, Nguyen Alan, Nieman Elizabeth, Meadows Norin, Gentry Tracy, Foster Sandra J, Balber Andrew E, Fiordalisi Michelle, Hinson James M. Development of a flow cytometry-based pulse-width assay for detection of aggregates in cellular therapeutics to be infused by catheter.. Cytotherapy. 2014. DOI: 10.1016/j.jcyt.2014.05.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.