Cryoballoon pulmonary vein isolation supported by intracardiac echocardiography: integration of a nonfluoroscopic imaging technique in atrial fibrillation ablation.

Nölker Georg, Heintze Johannes, Gutleben Klaus-Jürgen, Muntean Bogdan, Pütz Vanessa, Yalda Ameera, Vogt Jürgen, Horstkotte Dieter

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

cryoballoon ablation has been adopted for pulmonary vein (PV) isolation (PVI) in many centers. Complete occlusion of PV by an adequately sized balloon is crucial for effectiveness of cryoenergy delivery. The aim of this study was to evaluate intracardiac echocardiography (ICE) as an alternative imaging technique compared to angiographic imaging in cryoballoon PVI.

METHODS AND

RESULTS

a total of 75 PVs were treated in 22 patients (61 ± 13 years, 17 male) undergoing PV cryoballoon ablation for drug refractory paroxysmal atrial fibrillation. Decision for an adequate balloon size was based on diameters of the PV antra assessed by ICE and PV angiography. Per PV 2.4 ± 0.4 cryoenergy pulses were applied. Decision for the balloon size was similar either based upon angiography or on ICE. A single 23 or 28 mm balloon was chosen in 10 and 3 patients, respectively. Two different sized balloons were used in 9 patients. PVI was evaluated after 2 cryoenergy applications. Total occlusion of the PV confirmed by ICE color flow Doppler (CFD) during ablation predicted successful PVI in 70 of 75 (93%) and unsuccessful PVI in 8 of 8 (100%). PV flow registered by pulsed wave Doppler at the PV ostium pre- and postablation was 0.48 ± 0.10 and 0.51 ± 0.12 m/s, respectively (n.s.). PVI was finally confirmed by entrance block in all PVs. No procedural complications occurred.

CONCLUSIONS

ICE is a feasible novel imaging technique in cryoballoon ablation procedures. It allows decision for adequate balloon size, exact balloon placement, prediction of acute ablation success, and excludes acute narrowing of PV ostia.

Для цитирования: Nölker Georg, Heintze Johannes, Gutleben Klaus-Jürgen, Muntean Bogdan, Pütz Vanessa, Yalda Ameera, Vogt Jürgen, Horstkotte Dieter. Cryoballoon pulmonary vein isolation supported by intracardiac echocardiography: integration of a nonfluoroscopic imaging technique in atrial fibrillation ablation.. Journal of cardiovascular electrophysiology. 2010. DOI: 10.1111/j.1540-8167.2010.01813.x.