DOI 10.1177/152660280200900104

Медицинская библиотека · Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2002

"Seat belt and air bag" technique for cerebral protection during carotid stenting.

Parodi Juan C, Schönholz Claudio, Ferreira L Mariano, Mendaro Esteban, Ohki Takao

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To present a 2-device technique for cerebral protection during carotid stenting of the internal carotid artery (ICA) in patients with high-grade lesions, contralateral occlusion, and/or an aberrant or nonfunctioning circle of Willis.

TECHNIQUE: A reverse flow system (Parodi Anti-Embolism System [PAES]) is first placed with a 260-cm exchange wire in the common carotid artery 3 cm below the carotid bifurcation. Flow reversal is obtained by inflating the balloons in the external and common carotid arteries. Via an external connector, a guidewire and E-Trap filter are delivered to the distal ICA with active suction from a syringe on the PAES catheter. Once the filter is above the stenosis, the reversed flow is discontinued, and the procedure proceeds under cerebral protection with the filter.

CONCLUSIONS

The combination of 2 existing cerebral protection devices could, at least in theory, achieve what neither of them could independently. The "seat belt and air bag" approach may represent the ideal means of treating ICA stenoses in patients with a non-functioning circle of Willis.

Для цитирования: Parodi Juan C, Schönholz Claudio, Ferreira L Mariano, Mendaro Esteban, Ohki Takao. "Seat belt and air bag" technique for cerebral protection during carotid stenting.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2002. DOI: 10.1177/152660280200900104.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.